EXTRACTIONLESS HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF DICLOFENAC IN HUMAN PLASMA AND URINE

被引:35
作者
AVGERINOS, A [1 ]
KARIDAS, T [1 ]
MALAMATARIS, S [1 ]
机构
[1] UNIV THESSALONIKI,PHARMACEUT TECHNOL LAB,GR-54006 SALONIKA,GREECE
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1993年 / 619卷 / 02期
关键词
D O I
10.1016/0378-4347(93)80125-N
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An assay using reversed-phase high-performance liquid chromatography with ultraviolet detection, at 278 nm, was developed to measure diclofenac in human plasma and urine at concentrations suitable for biopharmaceutical studies. Indomethacin was used as internal standard and separation was performed at 40-degrees-C on a C18 Spherisorb column with acetonitrile-0.1 M sodium acetate (35:65, v/v) (pH 6.3) as mobile phase. The sample preparation is simple and rapid (extractionless), and the total run time is less than 5 min. The retention time is 2.8 min for diclofenac and 3.6 min for indomethacin. The detection limit is 0.2 ug/ml using a 20-mul loop.