RAPID AND HIGHLY SENSITIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF HISTAMINE AND 3-METHYLHISTAMINE IN BIOLOGICAL SAMPLES USING FLUORESCAMINE AS THE DERIVATIZING AGENT

被引:20
作者
VANHAASTER, CMCJ
ENGELS, W
LEMMENS, PJMR
HORNSTRA, G
VANDERVUSSE, GJ
机构
[1] UNIV LIMBURG,DEPT HUMAN BIOL,6200 MD MAASTRICHT,NETHERLANDS
[2] UNIV LIMBURG,DEPT PHYSIOL,6200 MD MAASTRICHT,NETHERLANDS
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1993年 / 617卷 / 02期
关键词
D O I
10.1016/0378-4347(93)80493-N
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly sensitive and rapid high-performance liquid chromatographic assay for the determination of histamine and 3-methyl-histamine in biological samples using 1-methylhistamine as the internal standard is described. Samples were purified and concentrated on cation-exchange columns and derivatized with fluorescamine. The lower detection limit was 20 pg on-column. Linearity was demonstrated up to 20 ng on-column. The samples could be derivatized simultaneously before injection and were stable for 7 days. The method was used for the determination of histamine and related compounds in coronary perfusates, extracts of homogenized rat hearts, and supernatants of stimulated peritoneal mast cells.
引用
收藏
页码:233 / 240
页数:8
相关论文
共 24 条
[1]   A DIRECT AND SENSITIVE DETERMINATION OF HISTAMINE IN ACID-DEPROTEINIZED BIOLOGICAL SAMPLES BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
ARAKAWA, Y ;
TACHIBANA, S .
ANALYTICAL BIOCHEMISTRY, 1986, 158 (01) :20-27
[2]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC TECHNIQUE FOR THE RAPID-DETERMINATION OF HISTAMINE IN BOTH THE PLASMA AND CELLULAR-COMPONENTS OF BLOOD [J].
ASHMORE, SP ;
THOMSON, AH ;
SIMPSON, H .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1989, 496 (02) :435-440
[3]   HISTAMINE ASSAY IN TEARS BY FLUORESCAMINE DERIVATIZATION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
BETTERO, A ;
ANGI, MR ;
MORO, F ;
BENASSI, CA .
JOURNAL OF CHROMATOGRAPHY, 1984, 310 (02) :390-395
[4]  
DAINAKA J, 1984, BIOCHEM PHARMACOL, V33, P1653
[5]  
FLANCBAUM L, 1989, CIRC SHOCK, V27, P155
[6]  
GALLI SJ, 1984, PROGR ALLERGY, V34, P90
[7]   FLUOROMETRIC-DETERMINATION OF HISTAMINE WITH OPT - OPTIMUM REACTION CONDITIONS AND TESTS OF IDENTITY [J].
HAKANSON, R ;
RONNBERG, AL ;
SJOLUND, K .
ANALYTICAL BIOCHEMISTRY, 1972, 47 (02) :356-+
[8]   REVERSED-PHASE ION-PAIR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC PROCEDURE FOR DETERMINATION OF HISTAMINE AND ITS METABOLITES IN RAT URINE [J].
HUI, JY ;
TAYLOR, SL .
JOURNAL OF CHROMATOGRAPHY, 1984, 312 (NOV) :443-449
[9]   SIMULTANEOUS DETERMINATION OF HISTIDINE-CONTAINING DIPEPTIDES, HISTAMINE, METHYLHISTAMINE AND HISTIDINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
KASZIBA, E ;
FLANCBAUM, L ;
FITZPATRICK, JC ;
SCHNEIDERMAN, J ;
FISHER, H .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 432 :315-320
[10]   INTERFERENCE WITH THE FLUOROMETRIC HISTAMINE ASSAY BY BIOGENIC-AMINES AND AMINO-ACIDS [J].
KOWNATZKI, E ;
GRUNINGER, G ;
FUHR, N .
PHARMACOLOGY, 1987, 34 (01) :17-24