NONPULSING REAGENT METERING FOR CONTINUOUS COLORIMETRIC DETECTION SYSTEMS

被引:11
作者
JOLLEY, RL
PITT, WW
SCOTT, CD
机构
[1] Oak Ridge National Laboratory, 2 2 Operated for the U. S. Atomic Energy Commission by Union Carbide Corporation., Oak Ridge
关键词
D O I
10.1016/0003-2697(69)90183-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A method has been developed for metering liquid reagents without variations in flow rate for continuous colorimetric detection systems. The reagent is forced from a storage chamber by hydraulic head or by gas overpressure, and flow control is achieved by resistance in the delivery line which may be either capillary tubing and/or a needle valve. Continuous flow of the reagent reaction mixture through the colorimeter is achieved by using a water aspirator. Such a system is now employed routinely in a continuous carbohydrate analyzer in which concentrated sulfuric acid and a 5% phenol solution are continously metered and mixed with the effluent from an anion-exchange resin column that separates the carbohydrates in physiological samples. © 1969.
引用
收藏
页码:300 / +
页数:1
相关论文
共 5 条
[1]  
JOLLEY RL, 1968, CLIN CHEM, V14, P538
[2]  
MOORE S, 1954, J BIOL CHEM, V211, P907
[3]  
NEWMAN DW, 1965, INSTRUMENTAL METH ED, P125
[4]   CONTINUOUS SEPARATION OF ION-EXCHANGE RESIN INTO SIZE FRACTIONS BY ELUTRIATION WITH WATER [J].
SCOTT, CD .
ANALYTICAL BIOCHEMISTRY, 1968, 24 (02) :292-&
[5]  
WEINSTEIN LH, 1965, INSTRUMENTAL METHODS, P125