UNUSUAL COMPLICATION OF TEMPORARY PACING WIRES IN CHILDREN

被引:24
作者
BOLTON, JWR [1 ]
MAYER, JE [1 ]
机构
[1] CHILDRENS HOSP MED CTR, DEPT CARDIAC SURG, BOSTON, MA 02115 USA
关键词
D O I
10.1016/0003-4975(92)91027-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
There have been numerous complications described concerning temporary epicardial pacing wires including infection, bleeding, and coronary artery graft disruption. We report a case in which the heart was ensnared by the pacing wire, resulting in cardiac decompensation during attempted removal. This complication may be avoided by assuring that a smaller loop of pacing wire remains in the pericardium.
引用
收藏
页码:769 / 770
页数:2
相关论文
共 1 条
[1]  
Del Nido P, 1989, J Card Surg, V4, P99, DOI 10.1111/j.1540-8191.1989.tb00262.x