CHARACTERIZATION OF THE CHEMICAL AND FUNCTIONAL STABILITY OF DEAE SEPHAROSE (R) FAST FLOW

被引:11
作者
ANDERSSON, M [1 ]
DREVIN, I [1 ]
JOHANSSON, BL [1 ]
机构
[1] PHARM BIOPROC TECHNOL AB, S-75182 UPPSALA, SWEDEN
关键词
D O I
10.1016/0032-9592(93)80038-I
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
The release of amines from the ion-exchange groups in DEAE Sepharose(R) Fast Flow has been studied under static and column conditions. The leakage compounds have been identified and quantified by gas chromatography-mass spectrometry. It was shown that the main leakage product under acidic (pH 1) and basic conditions (pH 14) was N,N,N',N'-tetraethylethylenediamine. Three other amines were also identified, namely N,N,N'-triethylethylenediamine, diethylaminoethanol and diethylamine. The leakage of amines from DEAE Sepharose Fast Flow treated at pH 1 or 14 for 672 h at 40-degrees-C corresponds to a reduction of only 1 % of the total ion-exchange capacity. The functional stability of DEAE Sepharose Fast Flow was also studied by separation of a protein mixture during repeated cleaning-in-place treatments with 0.10 M HCl or 1.0 M NaOH. The separation behaviour was unaltered after the gel had been treated for a total contact time of 672 h with 0.10 M HCl or 1.0 M NaOH. The clearance of ethanol from a DEAE Sepharose Fast Flow column stored in a 20 % ethanol aqueous solution was also studied.
引用
收藏
页码:223 / 230
页数:8
相关论文
共 11 条
[1]   SILICA DEXTRAN SORBENT COMPOSITES AND THEIR CLEANING IN PLACE [J].
BOSCHETTI, E ;
GIROT, P ;
GUERRIER, L .
JOURNAL OF CHROMATOGRAPHY, 1990, 523 :35-42
[2]   COLUMN PERFORMANCE OF Q-SEPHAROSE HP IN ANALYTICAL-SCALE AND PREPARATIVE-SCALE CHROMATOGRAPHY [J].
DREVIN, I ;
LARSSON, L ;
JOHANSSON, BL .
JOURNAL OF CHROMATOGRAPHY, 1989, 477 (02) :337-344
[3]   REGULATORY REQUIREMENTS FOR LICENSING MEDICINAL PRODUCTS OF BIOTECHNOLOGY [J].
DUNCAN, ME ;
CHARLESWORTH, FA ;
GRIFFIN, JP .
TRENDS IN BIOTECHNOLOGY, 1987, 5 (12) :325-328
[4]  
Gubensek F, 1968, J MACROMOL SCI CHEM, V2, P1045
[5]   HIGH-PERFORMANCE SIZE-EXCLUSION CHROMATOGRAPHY OF POLYPEPTIDES ON A TSK G2000SW COLUMN IN ACIDIC MOBILE PHASES [J].
IRVINE, GB .
JOURNAL OF CHROMATOGRAPHY, 1987, 404 (01) :215-222
[6]   DETERMINATION OF THE LEAKAGE FROM PHENYL-SEPHAROSE CL-4B, PHENYL-SEPHAROSE-FF AND PHENYL-SUPEROSE IN BULK AND COLUMN EXPERIMENTS [J].
JOHANSSON, BL ;
HELLBERG, U ;
WENNBERG, O .
JOURNAL OF CHROMATOGRAPHY, 1987, 403 :85-98
[7]  
KARLSSON E, 1989, PROTEIN PURIFICATION, P116
[8]  
KATO Y, 1982, J CHROMATOGR, V153, P219
[9]  
SOFER GK, 1989, PROCESS CHROMATOGRAP, P93
[10]  
WEININGER SJ, 1984, ORGANIC CHEM, P247