MONITORING NEUROMUSCULAR BLOCK BY ACCELEROMYOGRAPHY - COMPARISON OF THE MINI-ACCELOGRAPH WITH THE MYOGRAPH-2000

被引:52
作者
HARPER, NJN
MARTLEW, R
STRANG, T
WALLACE, M
机构
[1] Department of Anaesthesia, Manchester Royal Infirmary, Manchester M13 9WL, Oxford Road
[2] Department of Anaesthesia, Cairns Base Hospital, Queensland 4870, The Esplanade
关键词
MONITORING; NEUROMUSCULAR FUNCTION; NEUROMUSCULAR BLOCK; ATRACURIUM;
D O I
10.1093/bja/72.4.411
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The precision of the compact Mini-Accelograph (M-A) was compared with the Myograph 2000 (MYO). Neuromuscular block resulting from atracurium was measured simultaneously by the MYO and the M-A applied to contralateral thumbs. During onset, the M-A frequently underestimated the extent of block (maximal at approximately 50% twitch depression). The M-A control train-of-four (TOF) ratio was characteristically > 1.0 and remained greater than the MYO ratio during the onset of atracurium. During recovery, the difference between the MYO and the M-A was maximal at approximately 50% twitch depression, but the M-A frequently overestimated the extent of block. The mean differences between the MYO and the M-A were small in respect of the recovery index (RI) and the TOF. However, the limits of agreement were unacceptably wide for both TOF and RI. When the MYO TOF was 0.7, the corresponding M-A TOF varied between 0.4 and 1.0. The M-A was more susceptible to drift than the MYO.
引用
收藏
页码:411 / 414
页数:4
相关论文
共 5 条
  • [1] DETERMINANTS OF THE REVERSAL TIME OF COMPETITIVE NEUROMUSCULAR BLOCK BY ANTICHOLINESTERASES
    BEEMER, GH
    BJORKSTEN, AR
    DAWSON, PJ
    DAWSON, RJ
    HEENAN, PJ
    ROBERTSON, BA
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1991, 66 (04) : 469 - 475
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] BRAND JB, 1977, ANESTH ANALG, V56, P55
  • [4] THE ACCELERATION TRANSDUCER - AN ASSESSMENT OF ITS PRECISION IN COMPARISON WITH A FORCE DISPLACEMENT TRANSDUCER
    MAY, O
    NIELSEN, HK
    WERNER, MU
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1988, 32 (03) : 239 - 243
  • [5] VIBYMOGENSEN J, 1988, ACTA ANAESTH SCAND, V32, P45