WHICH DEPTH DOSE DATA SHOULD BE USED FOR DOSE PLANNING WHEN WEDGE FILTERS ARE USED TO MODIFY THE PHOTON-BEAM

被引:17
作者
KNOOS, T
WITTGREN, L
机构
[1] Dept. of Radiat. Phys., Lund Univ., Malmo
关键词
D O I
10.1088/0031-9155/36/2/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of beam data for open photon fields when calculating absorbed dose distributions for beams with wedge filters has been studied. The depth doses for beams with wedge filters are changed through beam hardening and the dose maximum can be shifted; both these changes result in errors in the final dose calculations of several per cent if open beam data are used. The errors are larger for 6 MV than for 18 MV x-rays. The depth of measurement for determining the wedge factor and the influence of other beam modifying devices are discussed. It is recommended that the reference depth be used instead of the dose maximum for these kinds of measurements since the influence of contaminating electrons in the beam will then be avoided and the wedge factor will be correct at a clinically relevant depth.
引用
收藏
页码:255 / 267
页数:13
相关论文
共 16 条
[1]   DETERMINATION OF EFFECTIVE BREMSSTRAHLUNG SPECTRA AND ELECTRON CONTAMINATION FOR PHOTON DOSE CALCULATIONS [J].
AHNESJO, A ;
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (10) :1451-1464
[2]   A PHOTON DOSE DISTRIBUTION MODEL EMPLOYING CONVOLUTION CALCULATIONS [J].
BOYER, A ;
MOK, E .
MEDICAL PHYSICS, 1985, 12 (02) :169-177
[3]   DOSIMETRIC PRECISION REQUIREMENTS IN RADIATION-THERAPY [J].
BRAHME, A .
ACTA RADIOLOGICA ONCOLOGY, 1984, 23 (05) :379-391
[4]  
BRAHME A, 1988, ACTA ONCOL S, V1
[5]   SCATTER DOSE FOR WEDGED FIELDS [J].
BROWN, LH ;
SIDDON, RL ;
BJARNGARD, BE .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (10) :1321-1326
[6]  
KNOOS T, 1987, 9TH P INT C US COMP
[7]   A DEPTH DEPENDENCE DETERMINATION OF THE WEDGE TRANSMISSION FACTOR FOR 4-10 MV PHOTON BEAMS [J].
MCCULLOUGH, EC ;
GORTNEY, J ;
BLACKWELL, CR .
MEDICAL PHYSICS, 1988, 15 (04) :621-623
[8]  
NILSSON B, 1985, THESIS U STOCKHOLM S
[9]  
NILSSON P, 1987, ESTRO LISBOA, P416
[10]  
Nordic Association of Clinical Physics, 1980, ACTA RADIOL ONCOL, V19, P55