OPTIMIZED INHALATION AEROSOLS .2. INERTIAL TESTING METHODS FOR PARTICLE-SIZE ANALYSIS OF PRESSURIZED INHALERS

被引:21
作者
PHILLIPS, EM
BYRON, PR
FULTS, K
HICKEY, AJ
机构
[1] VIRGINIA COMMONWEALTH UNIV,DEPT PHARM & PHARMACEUT,AEROSOL RES GRP,BOX 533 MCV STN,RICHMOND,VA 23298
[2] UNIV ILLINOIS,COLL PHARM M-C 880,CHICAGO,IL 60612
关键词
aerosols; inertial testing; inhalation; metered dose inhaler; particle size analysis; pressurized inhaler;
D O I
10.1023/A:1015973402316
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Pressurized metered dose inhaler (MDI) output from three different albuterol formulations was characterized using three inertial separation devices. Results were compared for the Delron six-stage cascade impactor (DCI6), the Andersen Mark II eight-stage impactor (ACI8), and Copley's twin-stage liquid impinger (LI). None of the devices tested in this study was ideal in all respects. All devices could differentiate between formulations in terms of respirable doses (albuterol amount with aerodynamic diameters <5.5 through 6.4 µm). Only the high-flow rate LI could differentiate among all three formulations when data were presented in terms of respirable percentage (RP) of drug collected. Values for RP were in excellent agreement for the independently calibrated impactors when the same evaporation chamber was used atop the impactors. The LI appeared to overestimate values for RP in vivo. Results are discussed in light of the debate surrounding the revision of USP aerosol testing requirements. Rigorous specifications for evaporation chambers and methodologies are necessary for meaningful inter- and intra-laboratory comparison of results when any of these devices are used. © 1990, Plenum Publishing Corporation. All rights reserved.
引用
收藏
页码:1228 / 1233
页数:6
相关论文
共 26 条