INVESTIGATION OF THE RESPONSE OF NAI(TL) SCINTILLATION DETECTORS USING AN X-RAY MONOCHROMATOR

被引:1
作者
SHERBINI, SS
SHAHABI, I
机构
[1] Nuclear Engineering Program, Department of Mechanical Engineering, Howard University, Washington, DC 20059, 2300 6th Street, N.W
来源
NUCLEAR INSTRUMENTS & METHODS | 1979年 / 164卷 / 03期
关键词
D O I
10.1016/0029-554X(79)90080-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The response of NaI(Tl) scintillation detectors to low energy photons was studied. The energy range extended from 25 keV to 90 keV. The variation of output pulse height as a function of photon energy was determined. Also the variation of resolution as a function of photon energy. A bent crystal spectrometer was used as the source of monoenergetic photons for the experiment. © 1979.
引用
收藏
页码:457 / 461
页数:5
相关论文
共 13 条
[1]   FLUORESCENT RESPONSE OF NAI(TL) AND CSI(TL) TO X RAYS AND GAMMA RAYS [J].
COLLINSON, AJL ;
HILL, R .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON, 1963, 81 (523) :883-&
[2]   NONLINEAR RESPONSE OF NAI(TL) TO PHOTONS [J].
ENGELKEMEIR, D .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1956, 27 (08) :589-591
[3]   STUDIES ON THE LINEARITY OF A SCINTILLATION GAMMA-RAY SPECTROMETER [J].
ERIKSEN, VO ;
JENSSEN, G .
PHYSICAL REVIEW, 1952, 85 (01) :150-150
[4]   THE RADIATIONS OF TH-231 (UY) [J].
FREEDMAN, MS ;
JAFFEY, AH ;
WAGNER, F ;
MAY, J .
PHYSICAL REVIEW, 1953, 89 (01) :302-309
[5]   A METHOD FOR GENERATING SINGLE GAMMA-RAY SHAPES FOR ANALYSIS OF SPECTRA [J].
HEATH, RL ;
HELMER, RG ;
SCHMITTROTH, LA ;
CAZIER, GA .
NUCLEAR INSTRUMENTS & METHODS, 1967, 47 (02) :281-+
[6]   VARIATION OF RESOLUTION WITH GAMMA ENERGY IN ACTIVATED ALKALI HALIDE CRYSTALS [J].
HILL, R ;
COLLINSON, AJ .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON, 1965, 85 (548P) :1067-+
[7]   THE NON-PROPORTIONAL RESPONSE OF NAI(T1) CRYSTALS TO GAMMA-RAYS [J].
IREDALE, P .
NUCLEAR INSTRUMENTS & METHODS, 1961, 11 (02) :336-339
[8]   THE NONPROPORTIONAL RESPONSE OF A NAI(TI) CRYSTAL TO DIFFRACTED X-RAYS [J].
JONES, TH .
NUCLEAR INSTRUMENTS & METHODS, 1962, 15 (01) :55-58
[9]  
LEDERER CM, 1968, TABLE ISOTOPES, P567
[10]   SHAPE OF PHOTOPEAK IN A NAI(TI) SCINTILLATION DETECTOR [J].
MUDHOLE, TS ;
UMAKANTHA, N .
NUCLEAR INSTRUMENTS & METHODS, 1974, 116 (02) :401-403