IMPLANT PROVISION OF KEY, PINCH AND POWER GRIPS IN A C6 TETRAPLEGIC

被引:7
作者
PERKINS, TA
BRINDLEY, GS
DONALDSON, ND
POLKEY, CE
RUSHTON, DN
机构
[1] MRC, NEUROL PROTHESES UNIT, LONDON SE5 8BB, ENGLAND
[2] MAUDSLEY HOSP & INST PSYCHIAT, NEUROSURG UNIT, LONDON SE5 8AZ, ENGLAND
关键词
FES; FINITE STATE CONTROL; GRIP SELECTION; HAND FUNCTION RESTORATION; IMPLANT; JOYSTICK; PORTABLE MICROCOMPUTER; TETRAPLEGIA;
D O I
10.1007/BF02524686
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
An 11-channel multiplexed stimulator of nerves and muscles in the left forearm was implanted for hand control in January 1986 in a 21 year old woman who, after sustaining a C6 spinal lesion 7 years earlier, had voluntary shoulder and elbow movement but paralysed hands, trunk and legs. The patient controls the stimulation via a microcomputer control box and an RF transdermal link. We have investigated the control of her stimulated hand with a joystick under her contralateral hand which she moves from the shoulder and elbow. Since 1986, we have tried a variety of joystick control schemes involving power and key grips. Currently, for grip adjustment, forward and backward joystick movements correspond to thumb extension and abduction, respectively giving in addition both finger and wrist extension, whereas right and left joystick movements yield fist closure and thumb opposition/adduction and flexion, respectively. Useful grasps are available by moving the joystick forward and then left (key grip), by moving the joystick backward and left (pinch grip), or by moving the joystick back and right (power grip). Thus, three distinct grips may be selected using these three quadrants of joystick movement. An additional control mode was found to be desirable to augment the patient's limited voluntary wrist positioning and provide wrist stability while adjusting finger grip.
引用
收藏
页码:367 / 372
页数:6
相关论文
共 11 条
[1]  
ANDREWS BJ, 1989, AUTOMEDICA, V11, P175
[2]  
BRINDLEY GS, 1989, JAN P BIOL ENG SOC C, P41
[3]  
Donaldson N., 1986, 2ND P VIENN INT WORK, P197
[4]  
DONALDSON N, 1990, ADV EXTERNAL CONTROL, P345
[5]  
DONALDSON ND, 1986, MED BIOL ENG COMPUT, V24, P50, DOI 10.1007/BF02441605
[7]   EVALUATION OF SHOULDER MOVEMENT AS A COMMAND CONTROL SOURCE [J].
JOHNSON, MW ;
PECKHAM, PH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (09) :876-885
[8]  
KEITH MW, 1988, CLIN ORTHOP RELAT R, P25
[9]   FNS OF THE UPPER LIMB - TARGETING THE FOREARM MUSCLES FOR SURFACE STIMULATION [J].
NATHAN, RH .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (03) :249-256
[10]   CONTROLLED PREHENSION AND RELEASE IN THE C5-QUADRIPLEGIC ELICITED BY FUNCTIONAL ELECTRICAL-STIMULATION OF THE PARALYZED FOREARM MUSCULATURE [J].
PECKHAM, PH ;
MORTIMER, JT ;
MARSOLAIS, EB .
ANNALS OF BIOMEDICAL ENGINEERING, 1980, 8 (4-6) :369-388