SUPERCRITICAL-FLUID EXTRACTION OF ORGANOCHLORINE PESTICIDES FROM AN AQUEOUS MATRIX

被引:41
作者
BARNABAS, IJ
DEAN, JR
HITCHEN, SM
OWEN, SP
机构
[1] UNIV NORTHUMBRIA,DEPT CHEM & LIFE SCI,NEWCASTLE TYNE NE1 8ST,ENGLAND
[2] ANALYT & ENVIRONM SERV,WALLSEND NE28 0QD,TYNE & WEAR,ENGLAND
关键词
D O I
10.1016/0021-9673(94)85060-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Supercritical fluid extraction (SFE) conditions were optimised for the removal of organochlorine pesticides (OCPs) from water. OCPs were collected and extracted from solid-phase extraction disks (Empore) and also directly from a water sample using a modified extraction cell. High recoveries (>90%) were obtained for two of the three OCPs with Empore disks. Despite the good solubility of OCPs in pure CO2, the analyte recoveries decreased when they were extracted directly from water. Three different flow-rates were used in the direct SFE with no apparent change in recovery, indicating that extraction was diffusion-controlled. The effect of increasing the ionic strength of the aqueous sample on analyte recovery was investigated.