DETERMINATION OF URANIUM IN ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:94
作者
BOOMER, DW [1 ]
POWELL, MJ [1 ]
机构
[1] MINIST ENVIRONM,REXDALE M9W 5L1,ONTARIO,CANADA
关键词
MASS SPECTROMETERS - SPECTROSCOPY - Applications - URANIUM AND ALLOYS - Spectroscopic Analysis;
D O I
10.1021/ac00150a019
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An analytical technique using inductively coupled plasma mass spectrometry has been developed to estimate the concentration of uranium in a variety of environmental samples. The lower limit for quantitation is 0. 1 ng/mL. Calibration is linear from the low limit to 1000 ng/mL. Precision, accuracy, and a quality control protocol have been established. A comparison with the conventional fluorometric method has been performed.