CLINICAL UTILITY OF TELEMETERED INTRACARDIAC ELECTROGRAMS IN DIAGNOSING A DESIGN DEPENDENT LEAD MALFUNCTION

被引:8
作者
SARMIENTO, JJ [1 ]
机构
[1] ST FRANCIS MED CTR,PEORIA,IL
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 02期
关键词
false inhibition; intracardiac electrograms; lead malfunction; spurious signals;
D O I
10.1111/j.1540-8159.1990.tb05069.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inhibition of pacemakers due to false signals from malfunctioning pacing leads has been previously reported. Three cases are reported in which inappropriale pauses were observed shortly following implantation. In all three cases, active fixation leads with an electronically active screw and ring tip electrode were used. All leads were manufactured by Oscor Medical, St. Petersburg, Florida. In case one, models PY‐61 (ventricle) and PY‐51 (atrium) were used. In case two, models PY‐61 were used in the ventricle and the atrium. In case three, models PY‐52 were used both in the atrium and ventricle. Thresholds at the time of implantation were acceptable. In cases one and two, inappropriate pauses were noted following, paced ventricular beats. In case three, inappropriate inhibition of the ventricular output was noted intermittently after some of the paced atrial beats. Spurious signals were identified as the cause of the apparent oversensing problem by using electrocardiograms with annotated telemetry information and noninvasively telemetered intracardiac electrograms. The amplitude of the spurious signals varied between 2 and 14 mV. In all cases, the problems resolved with time. A mechanism for the generation of false transients is proposed and the value of telemetered intracardiac electrograms is discussed. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:188 / 195
页数:8
相关论文
共 13 条
[1]  
Barold S S, 1985, Cardiol Clin, V3, P565
[2]  
BAROLD SS, 1983, EUR HEART J, V4, P724
[3]  
CLARKE M, 1987, J ELECTROPHYSIOL, V1, P388
[4]   ELECTRODE CAUSATION OF PACEMAKER INHIBITION [J].
FURMAN, S ;
DECAPRIO, V .
CHEST, 1977, 72 (01) :117-119
[5]   COMPARISON OF ACTIVE AND PASSIVE ADHERING LEADS FOR ENDOCARDIAL PACING [J].
FURMAN, S ;
PANNIZZO, F ;
CAMPO, I .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (04) :417-427
[6]   MECHANISM FOR FALSE INHIBITION OF DEMAND PACEMAKERS [J].
LASSETER, KC ;
BUCHANAN, JW ;
YOSHONIS, KF .
CIRCULATION, 1970, 42 (06) :1093-&
[7]   CLINICAL BENEFITS OF TELEMETERED ELECTROGRAMS IN ASSESSMENT OF DDD FUNCTION [J].
LEVINE, PA ;
SHOLDER, J ;
DUNCAN, JL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (06) :1170-1177
[8]  
Levine PA, 1987, J ELECTROPHYSIOL, V1, P144
[9]  
Levine PA, 1987, J ELECTROPHYSIOL, V1, P404
[10]   METHODS FOR RECORDING INTERMITTENT CONTACT SIGNALS IN DEMAND PACEMAKER ARRHYTHMIAS (11 CASES) [J].
MUGICA, J ;
DUCONGE, R ;
DUBOS, M ;
LAZARUS, B .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1978, 1 (02) :222-230