A GEOMETRICALLY ACCURATE VASCULAR PHANTOM FOR COMPARATIVE-STUDIES OF X-RAY, ULTRASOUND, AND MAGNETIC-RESONANCE VASCULAR IMAGING - CONSTRUCTION AND GEOMETRICAL VERIFICATION

被引:51
作者
FRAYNE, R
GOWMAN, LM
RICKEY, DW
HOLDSWORTH, DW
PICOT, PA
DRANGOVA, M
CHU, KC
CALDWELL, CB
FENSTER, A
RUTT, BK
机构
[1] JOHN P ROBARTS RES INST, TOM LAWSON FAMILY IMAGING RES LABS, LONDON N6A 5K8, ON, CANADA
[2] UNIV WESTERN ONTARIO, DEPT DIAGNOST RADIOL, LONDON N6A 3K7, ONTARIO, CANADA
[3] UNIV TORONTO, DEPT MED BIOPHYS, TORONTO M5S 1A1, ONTARIO, CANADA
[4] UNIV TORONTO, DEPT RADIOL, TORONTO M5S 1A1, ONTARIO, CANADA
关键词
VASCULAR IMAGING; VASCULAR MODELING; HEMODYNAMIC STUDIES; MULTIMODALITY PHANTOMS;
D O I
10.1118/1.597141
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A technique for producing accurate models of vascular segments for use in experiments that assess vessel geometry and flow has been developed and evaluated. The models are compatible with x-ray, ultrasound, and magnetic resonance (MR) imaging systems. In this paper, a model of the human carotid artery bifurcation, is evaluated that has been built using this technique. The phantom consists of a thin-walled polyester-resin replica of the bifurcation through which a blood-mimicking fluid may be circulated. The phantom is surrounded by an agar tissue-mimicking material and a series of fiducial markers. The blood- and tissue-mimicking materials have x-ray, ultrasound, and MR properties similar to blood and tissue; fiducial markers provide a means of aligning images acquired by different modalities. The root-mean-square difference between the inner wall geometry of the constructed model and the desired dimensions was 0.33 mm. Static images were successfully acquired using x-ray, ultrasound, and MR imaging systems, and are free of significant artifacts. Flow images acquired with ultrasound and MR agree qualitatively with each other, and with previously published flow pattems. Volume-flow measurements obtained with ultrasound and MR were within 4.4% of the actual values.
引用
收藏
页码:415 / 425
页数:11
相关论文
共 32 条
  • [1] STEADY FLOW IN A MODEL OF THE HUMAN CAROTID BIFURCATION .2. LASER-DOPPLER ANEMOMETER MEASUREMENTS
    BHARADVAJ, BK
    MABON, RF
    GIDDENS, DP
    [J]. JOURNAL OF BIOMECHANICS, 1982, 15 (05) : 363 - 378
  • [2] STEADY FLOW IN A MODEL OF THE HUMAN CAROTID BIFURCATION .1. FLOW VISUALIZATION
    BHARADVAJ, BK
    MABON, RF
    GIDDENS, DP
    [J]. JOURNAL OF BIOMECHANICS, 1982, 15 (05) : 349 - 362
  • [3] BOTTOMLEY PA, 1984, MED PHYS, V11, P425, DOI 10.1118/1.595535
  • [4] CARO CG, 1978, MECHANICS CIRCULATIO, P153
  • [5] GEOMETRIC MODELING OF THE CAROTID BIFURCATION IN HUMANS - IMPLICATIONS IN ULTRASONIC DOPPLER AND RADIOLOGIC INVESTIGATIONS
    FORSTER, FK
    CHIKOS, PM
    FRAZIER, JS
    [J]. JOURNAL OF CLINICAL ULTRASOUND, 1985, 13 (06) : 385 - 390
  • [6] COMPUTER-CONTROLLED FLOW SIMULATOR FOR MR FLOW STUDIES
    FRAYNE, R
    HOLDSWORTH, DW
    GOWMAN, LM
    RICKEY, DW
    DRANGOVA, M
    FENSTER, A
    RUTT, BK
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (05): : 605 - 612
  • [8] COMPREHENSIVE COMPILATION OF EMPIRICAL ULTRASONIC PROPERTIES OF MAMMALIAN-TISSUES
    GOSS, SA
    JOHNSTON, RL
    DUNN, F
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1978, 64 (02) : 423 - 457
  • [9] MULTIMODALITY IMAGE INTEGRATION FOR STEREOTAXIC SURGICAL PLANNING
    HENRI, CJ
    COLLINS, DL
    PETERS, TM
    [J]. MEDICAL PHYSICS, 1991, 18 (02) : 167 - 177
  • [10] A TIME-DELAY INTEGRATION CHARGE-COUPLED DEVICE CAMERA FOR SLOT-SCANNED DIGITAL RADIOGRAPHY
    HOLDSWORTH, DW
    GERSON, RK
    FENSTER, A
    [J]. MEDICAL PHYSICS, 1990, 17 (05) : 876 - 886