AMBULATORY MONITORING OF INDIRECT BEAT-TO-BEAT ARTERIAL-PRESSURE IN HUMAN FINGERS BY A VOLUME-COMPENSATION METHOD

被引:17
作者
KAWARADA, A [1 ]
SHIMAZU, H [1 ]
ITO, H [1 ]
YAMAKOSHI, K [1 ]
机构
[1] HOKKAIDO UNIV,APPL ELECT RES INST,SAPPORO,HOKKAIDO 060,JAPAN
关键词
AMBULATORY MEASUREMENT; DIGITAL ARTERY; INDIRECT SPHYGMOMANOMETRY; INSTANTANEOUS ARTERIAL PRESSURE;
D O I
10.1007/BF02446296
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Using a volume-compensation technique, a portable device has been designed for the indirect measurement of beat-to-beat arterial pressure and its waveforms in the basal phalanx of fingers of ambulatory subjects. The device consists of (1) a transmission infra-red photoelectric plethysmograph (TIPP) to detect the variation of arterial volume, (2) a pneumatic cuff with an actuator, (3) a servosystem to control the cuff pressure, and (4) a stereo cassette tape recorder. Arterial pressure was determined from the cuff pressure which was controlled by the servosystem so as to maintain the arterial volume constant at the 'vascular unloading' state. This device is equipped with a compensator for any hydrostatic pressure difference between the heart and finger. Thus, the blood pressure at heart level can be obtained for any finger height. The total weight of the device was 1.6kg. Blood pressure changes during walking, jogging, jumping, and exercises such as side-stepping, Master's two-step test and car driving, have been successfully recorded.
引用
收藏
页码:55 / 62
页数:8
相关论文
共 10 条
[1]   EFFECTS OF PERIPHERAL VASOCONSTRICTION ON THE BLOOD-PRESSURE IN THE FINGER, MEASURED CONTINUOUSLY BY A NEW NONINVASIVE METHOD (THE FINAPRES) [J].
DORLAS, JC ;
NIJBOER, JA ;
BUTIJN, WT ;
VANDERHOEVEN, GMA ;
SETTELS, JJ ;
WESSELING, KH .
ANESTHESIOLOGY, 1985, 62 (03) :342-345
[2]   PORTABLE BLOOD PRESSURE RECORDER ACCURACY AND PRELIMINARY USE IN EVALUATING INTRADAILY VARIATIONS IN PRESSURE [J].
HINMAN, AT ;
ENGEL, BT ;
BICKFORD, AF .
AMERICAN HEART JOURNAL, 1962, 63 (05) :663-&
[3]   AMBULATORY BLOOD-PRESSURE MONITORING - RECENT ADVANCES AND CLINICAL-APPLICATIONS [J].
HORAN, MJ ;
PADGETT, NE ;
KENNEDY, HL .
AMERICAN HEART JOURNAL, 1981, 101 (06) :843-848
[4]  
Penaz J., 1973, 10TH INT C MED BIOL, P104
[5]  
POSEY J A, 1969, Cardiovascular Research Center Bulletin (Houston), V8, P15
[6]  
Wesseling KH, 1982, FUNKT BIOL MED, V1, P245
[7]   NEW OSCILLOMETRIC METHOD FOR INDIRECT MEASUREMENT OF SYSTOLIC AND MEAN ARTERIAL-PRESSURE IN THE HUMAN FINGER .1. MODEL EXPERIMENT [J].
YAMAKOSHI, K ;
SHIMAZU, H ;
SHIBATA, M ;
KAMIYA, A .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1982, 20 (03) :307-313
[8]   INDIRECT MEASUREMENT OF INSTANTANEOUS ARTERIAL BLOOD-PRESSURE IN THE HUMAN FINGER BY THE VASCULAR UNLOADING TECHNIQUE [J].
YAMAKOSHI, K ;
SHIMAZU, H ;
TOGAWA, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (03) :150-155
[9]   NON-INVASIVE AUTOMATIC MONITORING OF INSTANTANEOUS ARTERIAL BLOOD-PRESSURE USING THE VASCULAR UNLOADING TECHNIQUE [J].
YAMAKOSHI, K ;
KAMIYA, A ;
SHIMAZU, H ;
ITO, H ;
TOGAWA, T .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1983, 21 (05) :557-565
[10]   INDIRECT MEASUREMENT OF INSTANTANEOUS ARTERIAL BLOOD-PRESSURE IN THE RAT [J].
YAMAKOSHI, KI ;
SHIMAZU, H ;
TOGAWA, T .
AMERICAN JOURNAL OF PHYSIOLOGY, 1979, 237 (05) :H632-H637