ACCURACY AND PRECISION OF GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY FOR STABLE CARBON-ISOTOPE RATIO MEASUREMENTS

被引:26
作者
WONG, WW
HACHEY, DL
ZHANG, S
CLARKE, LL
机构
[1] USDA, ARS Children's Nutrition Research Center, Department of Pediatrics, Baylor College of Medicine, Houston, Texas, 77030
关键词
D O I
10.1002/rcm.1290091107
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We evaluated the precision, accuracy and linearity of three commercial gas chromatography/combustion isotope ratio mass spectrometry (GC/C-IRMS) instruments in regard to stable carbon isotope ratio measurements of fatty acids with C-13 enrichment levels ranging from natural abundance to 924 parts per thousand, in comparison to measurements by conventional gas isotope ratio mass spectrometry (GIRMS). Memory effect was also evaluated with a fatty acid mixture containing methyl myristate and methyl stearate at natural abundance levels of C-13 but methyl palmitate at C-13 enrichment levels of 460 or 924 parts per thousand. At natural abundance, the delta(13)C values of fatty acids measured by the GC/C-IRMS instruments were reproducible to within 0.34 parts per thousand (range: 0.11 parts per thousand to 0.32 parts per thousand and accurate to -0.38 +/- 0.40 parts per thousand (mean difference +/- standard deviation; range: -0.03 +/- 0.25 parts per thousand to -0.84 +/- 0.39 parts per thousand). At enrichment levels of C-13 (23 to 994 parts per thousand, the delta(13)C values were reproducible to within 1.48 parts per thousand (range 0.20 parts per thousand to 1.00 parts per thousand and accurate to -1.11+/-3.16 parts per thousand (range: -1.02 +/- 3.23 parts per thousand to +4.56 +/- 13.42 parts per thousand). Stable carbon isotope ratio measurements by all three GC/C-IRMS instruments were found to be linear from -30 to 924 parts per thousand (r(2) > 0.9999; p less than or equal to 0.004) in comparison to those measured by GIRMS. Significant memory effect was found in only one system. Based on our data, the GC/C-IRMS instruments can provide linear, precise and accurate stable carbon isotope ratio measurements and require only pico- and nanogram quantities of samples.
引用
收藏
页码:1007 / 1011
页数:5
相关论文
共 17 条
[1]   GAS-CHROMATOGRAPHY - STABLE ISOTOPE RATIO ANALYSIS AT NATURAL ABUNDANCE LEVELS [J].
BARRIE, A ;
BRICOUT, J ;
KOZIET, J .
BIOMEDICAL MASS SPECTROMETRY, 1984, 11 (11) :583-588
[2]   GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY ANALYSIS OF URINARY STEROIDS TO DETECT MISUSE OF TESTOSTERONE IN SPORT [J].
BECCHI, M ;
AGUILERA, R ;
FARIZON, Y ;
FLAMENT, MM ;
CASABIANCA, H ;
JAMES, P .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1994, 8 (04) :304-308
[3]   COMPARISON OF QUARTZ AND PYREX TUBES FOR COMBUSTION OF ORGANIC-SAMPLES FOR STABLE CARBON ISOTOPE ANALYSIS [J].
BOUTTON, TW ;
WONG, WW ;
HACHEY, DL ;
LEE, LS ;
CABRERA, MP ;
KLEIN, PD .
ANALYTICAL CHEMISTRY, 1983, 55 (11) :1832-1833
[4]   PREPARATION AND STABLE ISOTOPE DETERMINATION OF NBS-16 AND NBS-17 CARBON-DIOXIDE REFERENCE SAMPLES [J].
COPLEN, TB ;
KENDALL, C .
ANALYTICAL CHEMISTRY, 1982, 54 (14) :2611-2612
[5]   COMPARISON OF STABLE ISOTOPE REFERENCE SAMPLES [J].
COPLEN, TB ;
KENDALL, C ;
HOPPLE, J .
NATURE, 1983, 302 (5905) :236-238
[6]   THE GEOCHEMISTRY OF THE STABLE CARBON ISOTOPES [J].
CRAIG, H .
GEOCHIMICA ET COSMOCHIMICA ACTA, 1953, 3 (2-3) :53-92
[8]  
Glantz S., 1981, PRIMER BIOSTATISTICS
[9]   HIGH-SENSITIVITY TRACER DETECTION USING HIGH-PRECISION GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY AND HIGHLY ENRICHED [U-C-13]-LABELED PRECURSORS [J].
GOODMAN, KJ ;
BRENNA, JT .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1088-1095
[10]   COMPOUND-SPECIFIC CARBON-ISOTOPE RATIO DETERMINATION OF ENRICHED CHOLESTEROL [J].
GUO, ZK ;
LUKE, AH ;
LEE, WP ;
SCHOELLER, D .
ANALYTICAL CHEMISTRY, 1993, 65 (15) :1954-1959