SEMIAUTOMATED ENZYMIC MICRO METHODS FOR BLOOD GLUCOSE AND LACTIC ACID ON A SINGLE FILTRATE

被引:33
作者
ASROW, G
机构
[1] University of Illinois, College of Medicine, Department of Pediatrics, Chicago, IL 60612
关键词
D O I
10.1016/0003-2697(69)90164-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Methods have been modified and adapted to permit enzymic automated micro analysis of glucose on 0.021 ml of blood, and of lactic acid on 0.0115 ml of blood. By means of a single 1:20 Somogyi filtrate with a final volume of 1-1.4 ml prepared from 0.1-0.15 ml blood, both determinations can be accomplished individually on the autoanalyzer. The range of concentrations of each method has been extended so that glucose may be determined between 5 and 200 mg 100 ml (or higher depending on dilution) and lactic acid between 5 and 125 mg 100 ml. The methods are sensitive, reproducible, and stable. The recoveries and the comparison studies with manual methods are well within the limits of clinical investigation procdures. © 1969.
引用
收藏
页码:130 / &
相关论文
共 13 条
[1]   NEW AUTOMATED METHOD FOR MEASURING GLUCOSE BY GLUCOSE OXIDASE [J].
CRAMP, DG .
JOURNAL OF CLINICAL PATHOLOGY, 1967, 20 (06) :910-&
[3]   AN AUTOMATED MICRO DETERMINATION OF BLOOD-GLUCOSE WITH AUTOANALYZER [J].
FAULKNER, DE .
ANALYST, 1965, 90 (1077) :736-&
[4]   A RAPID ENZYMIC DETERMINATION OF L(+)-LACTIC ACID [J].
FRIEDLAND, I ;
DIETRICH, LS .
ANALYTICAL BIOCHEMISTRY, 1961, 2 (04) :390-&
[5]  
GETCHELL G, 1964, CLIN CHEM, V10, P540
[6]  
HILL JB, 1961, J LAB CLIN MED, V57, P970
[7]   AUTOMATED LACTIC ACID DETERMINATION IN SERUM AND TISSUE EXTRACTS [J].
HOCHELLA, NJ ;
WEINHOUSE, S .
ANALYTICAL BIOCHEMISTRY, 1965, 10 (02) :304-+
[9]  
ROBIN M, 1965, CLIN CHEM, V11, P840
[10]  
SAIFER A, 1958, J LAB CLIN MED, V51, P448