An equation for the random noise of self-scanned silicon photodiode arrays is developed which includes significant terms not previously considered. It is shown that, including these terms, the new 2.5-mm-high self-scanned arrays are only slightly noisier than older arrays; but because of their much greater aperture, they can detect much smaller light fluxes than the older arrays even when used with simple, inexpensive amplifiers.
引用
收藏
页码:730 / 732
页数:3
相关论文
共 13 条
[1]  
BUSS RR, 1976, SOLID STATE IMAGING
[2]  
DESSEY RE, 1976, J CHROMATOGR SCI, V14
[3]   CHARACTERISTICS OF PHOTODIODE ARRAYS FOR SPECTROCHEMICAL MEASUREMENTS [J].
HORLICK, G .
APPLIED SPECTROSCOPY, 1976, 30 (02) :113-123
[4]   SOLAR MAGNETOGRAPH EMPLOYING INTEGRATED DIODE ARRAYS [J].
LIVINGSTON, WC ;
HARVEY, J ;
SLAUGHTER, C ;
TRUMBO, D .
APPLIED OPTICS, 1976, 15 (01) :40-52
[5]  
MCNALL JF, 1973, P S I ASTRON SPACE S
[6]   RAPID SCANNING DIODE-ARRAY AS A MULTI-WAVELENGTH DETECTOR IN LIQUID-CHROMATOGRAPHY [J].
MILANO, MJ ;
LAM, S ;
GRUSHKA, E .
JOURNAL OF CHROMATOGRAPHY, 1976, 125 (01) :315-326
[7]   DIODE-ARRAY SPECTROMETER FOR SIMULTANEOUS DETERMINATION OF HEMOGLOBINS IN WHOLE-BLOOD [J].
MILANO, MJ ;
KIM, KY .
ANALYTICAL CHEMISTRY, 1977, 49 (04) :555-559
[8]   DIODE-ARRAY DETECTOR IN LIQUID-CHROMATOGRAPHY .2. ENHANCED SENSITIVITY VIA FIRST DERIVATIVE (DA-DLAMBDA) CHROMATOGRAMS [J].
MILANO, MJ ;
GRUSHKA, E .
JOURNAL OF CHROMATOGRAPHY, 1977, 133 (02) :352-354
[9]  
PERKO L, 1977, P SOC PHOTOOPT INSTR, V116, P56
[10]  
SNOW EH, 1978, P ELECTRO OPTIC LASE