A BIOMECHANICAL COMPARISON OF GARDNER-WELLS TONGS AND HALO DEVICE USED FOR CERVICAL-SPINE TRACTION

被引:7
作者
LERMAN, JA [1 ]
HAYNES, RJ [1 ]
KOENEMAN, EJ [1 ]
KOENEMAN, JB [1 ]
WONG, BB [1 ]
机构
[1] HARRINGTON ARTHRITIS RES CTR,PHOENIX,AZ
关键词
Cervical spine; Gardner-Wells tongs; Halo; Traction;
D O I
10.1097/00007632-199411000-00006
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Summary of Background Data. Unstable cervical spine fractures and dislocations are often reduced by the application of axial traction using a halo or Gardner-Wells tongs. Failure of tong or halo attachment can cause substantial morbidity and usually occurs at the pin-bone interface. Institutions commonly clean and reuse tongs. The effect of tong wear on pullout strength and the strength of the halo used as a traction device have not been documented. Methods. A skull model biomechanically similar to human calvarium was used to compare the axial pullout strengths of four sets of new tongs, three sets of rarely used tongs, and one set of heavily used tongs, as well as a standard four-pin halo. Results. The pullout strength of tongs tightened to the manufacturer's recommended level appeared to decrease with increased use. Measurement of the pin force generated by each set of tongs and of the spring constant of each spring, as well as inspection of the tongs after testing, suggested that the decrease in pullout strength may be partly attributable to spring and/or pin wear. The pullout strength of the halo or of the new or slightly used tongs but not the heavily used tongs exceeded the maximum weight used clinically in cervical spine traction. Conclusion. The data suggest that consideration be given to replacement or recalibration of heavily used tongs.
引用
收藏
页码:2403 / 2406
页数:4
相关论文
共 9 条
[1]  
Barr J.S., Krag M.H., Pierce D.S., Cranial traction and the halo orthosis, The Cervical Spine, pp. 299-311, (1989)
[2]  
Blumberg K.D., Catalano J.B., Coder J.M., Balderston R.A., The Pullout strength of titanium alloy MRI-compadble and stainless steel MRI-incompatible Gardner-Wells tongs, Spine, 18, pp. 1985-1996, (1993)
[3]  
Cotier J.M., Herbison G.J., Nasuti J.F., Ditunno J.F., An H., Wolff B.E., Closed reduction of traumatic cervical spine dislocation using traction weights up to 140pounds, Spine, 18, pp. 386-390, (1993)
[4]  
Gardner W.J., The principle of spring-loaded points for cervical traction, J Neurosurg, 39, pp. 543-544, (1973)
[5]  
Garfin S.R., Botte M.J., Waters R.L., Nickel V.L., Complications in the use of the halo fixation device, J Bone Joint Surg [Am], 68, pp. 320-325, (1986)
[6]  
Grundy D.J., Skull traction and its complications, Injury, 15, pp. 173-177, (1983)
[7]  
Krag M.H., Byrt W., Pope M., Pull-off strength of Gardner-Well tongs from cadaveric crania, Spine, 14, pp. 247-250, (1989)
[8]  
Rosenblatt S.L., Wong W.B., Koeneman J.B., Haynes R.J., The Halo Skeletal Fixator: A Design Modification
[9]  
Whitesides T.E., Mehserle W.L., Hutton W.C., The force exerted by the halo pin, Spine, 17, pp. 5413-5417, (1992)