MACROPOROUS POLYMERIC STATIONARY-PHASE ROD AS CONTINUOUS SEPARATION MEDIUM FOR REVERSED-PHASE CHROMATOGRAPHY

被引:277
作者
WANG, QC [1 ]
SVEC, F [1 ]
FRECHET, JMJ [1 ]
机构
[1] CORNELL UNIV,DEPT CHEM,BAKER LAB,ITHACA,NY 14853
关键词
D O I
10.1021/ac00065a013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A macroporous poly(styrene-co-divinylbenzene) rod has been prepared by a free-radical polymerization of a mixture containing monomers, initiator, and porogenic solvent in the confines of a chromatographic column and used for the first time in the very fast reversed-phase HPLC of proteins. Characterization of the pore structure of the continuous rod by mercury intrusion porosimetry revealed a large volume of pores with a diameter of about 1 mum to pores below 100 nm. Size exclusion chromatography and scanning electron microscopy confirmed the unusual pore size distribution. The presence of large pores make the rod easily permeable to eluents, and therefore, the back pressure of the rod column is modest even at high flow rates. The efficiency of the polymerized column is almost independent of the flow rate. The slope of the line showing capacity factor vs composition of the mobile phase was determined for several proteins, and a gradient for the separation of their mixtures was developed. Excellent separation was achieved even at a high flow rate of 25 mL/min as documented by the resolution data. Tripling the length of the column did not improve the column resolution in protein separation.
引用
收藏
页码:2243 / 2248
页数:6
相关论文
共 36 条
  • [1] CARRIER MEMBRANE AS A STATIONARY PHASE FOR AFFINITY-CHROMATOGRAPHY AND KINETIC-STUDIES OF MEMBRANE-BOUND ENZYMES
    ABOUREBYEH, H
    KORBER, F
    SCHUBERTREHBERG, K
    REUSCH, J
    JOSIC, D
    [J]. JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1991, 566 (02): : 341 - 350
  • [2] PERFUSION CHROMATOGRAPHY PACKING MATERIALS FOR PROTEINS AND PEPTIDES
    AFEYAN, NB
    FULTON, SP
    REGNIER, FE
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 544 (1-2): : 267 - 279
  • [3] FLOW-THROUGH PARTICLES FOR THE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION OF BIOMOLECULES - PERFUSION CHROMATOGRAPHY
    AFEYAN, NB
    GORDON, NF
    MAZSAROFF, I
    VARADY, L
    FULTON, SP
    YANG, YB
    REGNIER, FE
    [J]. JOURNAL OF CHROMATOGRAPHY, 1990, 519 (01): : 1 - 29
  • [4] CANTA G, 1992, SEP TECHNOL, V2, P62
  • [5] ALIGNED FIBER COLUMNS FOR SIZE-EXCLUSION CHROMATOGRAPHY
    CZOK, M
    GUIOCHON, G
    [J]. JOURNAL OF CHROMATOGRAPHY, 1990, 506 : 303 - 317
  • [6] SYNTHESIS AND CHARACTERIZATION OF 2-MU-M WIDE-PORE SILICA MICROSPHERES AS COLUMN PACKINGS FOR THE REVERSED-PHASE LIQUID-CHROMATOGRAPHY OF PEPTIDES AND PROTEINS
    DANIELSON, ND
    KIRKLAND, JJ
    [J]. ANALYTICAL CHEMISTRY, 1987, 59 (20) : 2501 - 2506
  • [7] VERY HIGH-SPEED SEPARATION OF PROTEINS WITH A 20-MU-M REVERSED-PHASE SORBENT
    FULTON, SP
    AFEYAN, NB
    GORDON, NF
    REGNIER, FE
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 547 (1-2): : 452 - 456
  • [8] Guyot A., 1982, PROG POLYM SCI, V8, P277, DOI [10.1016/0079-6700(82)90002-8., DOI 10.1016/0079-6700(82)90002-8]
  • [9] DETERMINATION OF PORE DISTRIBUTION OF MICROPOROUS SOLIDS BY EXCLUSION CHROMATOGRAPHY
    HALASZ, I
    VOGTEL, P
    GROH, R
    [J]. ZEITSCHRIFT FUR PHYSIKALISCHE CHEMIE-FRANKFURT, 1978, 112 (02): : 235 - 249
  • [10] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ON CONTINUOUS POLYMER BEDS
    HJERTEN, S
    LIAO, JL
    ZHANG, R
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 473 (01): : 273 - 275