FRETTING CORROSION STUDIES OF UNIVERSAL FEMORAL-HEAD PROSTHESES AND CONE TAPER SPIGOTS

被引:31
作者
FRICKER, DC
SHIVANATH, R
机构
[1] Howmedica International Inc., Limerick
关键词
Fretting corrosion; hip prostheses; modular heads;
D O I
10.1016/0142-9612(90)90064-W
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Short-duration, cyclically loaded, axial, fretting corrosion tests were designed and performed to compare the fretting behaviour of different metal Howmedica® universal modular heads connected to coated and uncoated metal cone taper spigots. Concurrent axial push-on and pull-off taper friction tests were also performed. There was no measurable fretting debris found in any test and SEM studies of the contact surfaces showed no evidence of fretting. It is concluded that no significant, long-term fretting corrosion of the Vitallium® Co-Cr-Mo heads on the Vitallium or uncoated Ti-6AI-4V spigots, nor of the Orthinox® stainless steel heads on the Orthinox spigots, is likely to occur in vivo. © 1990.
引用
收藏
页码:495 / 500
页数:6
相关论文
共 16 条
  • [1] ASGIAN C, 1984, 2ND T WORLD BIOM C W, P145
  • [2] FRETTING WEAR MECHANISMS AND THEIR EFFECTS ON FRETTING FATIGUE
    BERTHIER, Y
    COLOMBIE, C
    VINCENT, L
    GODET, M
    [J]. JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 1988, 110 (03): : 517 - 524
  • [3] FRETTING FATIGUE AND CONTACT CONDITIONS - A RATIONAL EXPLANATION OF PALLIATIVE BEHAVIOR
    CHIVERS, TC
    GORDELIER, SC
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 1985, 199 (04) : 325 - 337
  • [4] TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY
    DAVY, DT
    KOTZAR, GM
    BROWN, RH
    HEIPLE, KG
    GOLDBERG, VM
    HEIPLE, KG
    BERILLA, J
    BURSTEIN, AH
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) : 45 - 50
  • [5] DETERMINATION OF THE SERVICE LIFE OF A FRICTION LOCKING FIXATION ELEMENT FOR USE IN CERAMIC TOTAL HIPS
    DAWIHL, W
    DORRE, E
    ALTMEYER, G
    [J]. BIOMEDIZINISCHE TECHNIK, 1980, 25 (12): : 311 - 315
  • [6] Fessler H, 1989, Proc Inst Mech Eng H, V203, P1, DOI 10.1243/PIME_PROC_1989_203_003_01
  • [7] FESSLER H, 1989, J ENG MED, V203, P15
  • [8] GRAICHEN F, 1988, BIOMED TECH, V33, P305
  • [9] Johnson K. L., 1987, CONTACT MECH, P45
  • [10] KLEER G, 1985, BIOMECHANICS CURRENT, P353