EFFECT OF SECONDARY FLOW ON VISCOSITY MEASUREMENT USING CONE-AND-PLATE VISCOMETER

被引:44
作者
CHENG, DCH
机构
[1] Warren Spring Laboratory, Stevenage, Herts England
关键词
D O I
10.1016/0009-2509(68)80023-5
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The usefulness of the cone-and-plate viscometer is limited to rotational speeds low enough for the assumption of primary flow to be valid. Experimental data for Newtonian liquids show that a theoretical result of Walters and Waters [5] satisfactorily predicts the conditions at which the effect of secondary flow becomes appreciable. © 1968.