SCREENING FOR GESTATIONAL DIABETES-MELLITUS WITH A REFLECTANCE PHOTOMETER - ACCURACY AND PRECISION OF THE SINGLE-OPERATOR MODEL

被引:9
作者
MURPHY, NJ
MEYER, BA
OKELL, RT
HOGARD, ME
机构
[1] UNIV MISSOURI,ST LUKES HOSP,DEPT MED EDUC,KANSAS CITY,MO 64110
[2] UNIV MISSOURI,ST LUKES HOSP,DEPT MATERNAL FETAL MED,KANSAS CITY,MO 64110
[3] ST LUKES HOSP,DEPT LAB MED,KANSAS CITY,MO 64111
关键词
D O I
10.1097/00006250-199406000-00027
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective: To investigate the accuracy and precision of a recent-generation reflectance photometer, Accu-Chek III, for gestational diabetes mellitus screening with a single operator. Methods: Four hundred ten paired capillary and serum glucose samples collected prospectively were evaluated by a reflectance photometer and a glucose oxidase method (Beckman), respectively, during routine gestational diabetes mellitus screening with a single operator. Results: The mean Accu-Chek III result was 12% greater than serum glucose oxidase (115.5 +/- 1.4 and 103.4 +/- 1.3 mg/dL, respectively). A serum glucose level of at least 135 mg/dL by glucose oxidase produced sensitivity and specificity of 0.81 and 0.68, respectively. At 155 mg/dL, the Accu-Chek III sensitivity was 0.81 and specificity 0.74, and the method produced fewer false-positive results (37 of 410 [9.0%] versus 41 of 410 [10%]) than a 135-mg/dL serum cutoff. Adjusting the Accu-Chek III threshold by two times the coefficient of variation resulted in the same number of false positives as the Beckman technique at 135 mg/dL. The receiver operating characteristic curve cutoff for the venous value of 135 mg/dL was an Accu-Chek III value of 141 mg/dL. Linear regression analysis predicted y = 20.6 +/- 0.919x, r = 0.838. The mean coefficient of variation was 3.9%. Conclusions: The Accu-Chek III produced accurate and precise gestational diabetes mellitus screening results with a single operator. Because capillary blood is a combination of arterial and venous blood, an Accu-Chek III cutoff of 155 mg/dL may be more appropriate for screening, considering the 10-15% higher capillary glucose level in the fed state. An Accu-Chek III cutoff of 155 mg/dL produced a higher specificity than venous serum at 135 mg/dL, with no decrease in sensitivity.
引用
收藏
页码:1038 / 1042
页数:5
相关论文
共 25 条
[1]  
CARR S, 1989, OBSTET GYNECOL, V73, P727
[2]  
CENTOR RM, 1992, ROC ANAL
[3]   SHOULD THE 50-GRAM, ONE-HOUR PLASMA-GLUCOSE SCREENING-TEST FOR GESTATIONAL DIABETES BE ADMINISTERED IN THE FASTING OR FED STATE [J].
COUSTAN, DR ;
WIDNESS, JA ;
CARPENTER, MW ;
ROTONDO, L ;
PRATT, DC ;
OH, W .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1986, 154 (05) :1031-1035
[4]  
DAVIDSON MB, 1981, DIABETES MELLITUS DI, P3
[5]   COMPARISON OF 5 GLUCOSE METERS FOR SELF-MONITORING OF BLOOD-GLUCOSE BY DIABETIC-PATIENTS [J].
GIFFORDJORGENSEN, RA ;
BORCHERT, J ;
HASSANEIN, R ;
TILZER, L ;
EAKS, GA ;
MOORE, WV .
DIABETES CARE, 1986, 9 (01) :70-76
[6]   FOREARM GLUCOSE UPTAKE DURING ORAL GLUCOSE-TOLERANCE TEST IN NORMAL SUBJECTS [J].
JACKSON, RA ;
PETERS, N ;
ADVANI, U ;
PERRY, G ;
ROGERS, J ;
BROUGH, WH ;
PILKINGTON, TR .
DIABETES, 1973, 22 (06) :442-458
[7]   IDENTIFYING SOURCES OF ERROR IN SELF-MONITORING OF BLOOD-GLUCOSE [J].
JOVANOVICPETERSON, L ;
PETERSON, CM ;
DUDLEY, JD ;
KILO, C ;
ELLIS, B .
DIABETES CARE, 1988, 11 (10) :791-794
[8]   CAPILLARY BLOOD-GLUCOSE SCREENING FOR GESTATIONAL DIABETES - A PRELIMINARY INVESTIGATION [J].
LANDON, MB ;
CEMBROWSKI, GS ;
GABBE, SG .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1986, 155 (04) :717-721
[9]  
LATMAN N S, 1991, Biomedical Instrumentation and Technology, V25, P43
[10]   OBSERVATIONS ON BLOOD-GLUCOSE AND INSULIN DETERMINATIONS [J].
LIND, T ;
CHEYNE, GA ;
BROWN, G ;
GROOT, HAVD .
BRITISH MEDICAL JOURNAL, 1972, 3 (5822) :320-&