[1] OAK RIDGE NATL LAB,DIV CHEM,OAK RIDGE,TN 37831
来源:
JOURNAL OF THERMAL ANALYSIS
|
1992年
/
38卷
/
10期
关键词:
HEAT CAPACITY;
DATA ANALYSIS;
DIFFERENTIAL SCANNING CALORIMETER;
DSC;
SINGLE RUN;
D O I:
10.1007/BF02123979
中图分类号:
O414.1 [热力学];
学科分类号:
摘要:
An outline for the data analysis of single-run heat capacity measurments by dual sample DSC is presented with the following features: 1. Heat flow correction by subtracting the contribution due to the sample pan, including correction for mismatched pan masses. 2. Heat flow and temperature correction with a nonlinear temperature calibration, temperature lag correction, and heating rate correction. 3. Calculation of the cell constants for both cell positions and evaluation of the asymmetry factor between cell positions A and B. 4. Heat capacity calibration and calculation with slope and asymmetry correction. 5. Calculation of heat capacity for multiple runs. 6. Data curve fitting for heat capacity.