A CONTINUOUS-FLOW DENUDER FOR THE MEASUREMENT OF AMBIENT CONCENTRATIONS AND SURFACE-EXCHANGE FLUXES OF AMMONIA

被引:247
作者
WYERS, GP
OTIES, RP
SLANINA, J
机构
[1] Netherlands Energy Research Foundation ECN, 1755 ZG Petten
来源
ATMOSPHERIC ENVIRONMENT PART A-GENERAL TOPICS | 1993年 / 27卷 / 13期
关键词
ATMOSPHERIC POLLUTION; AMMONIA; DIFFUSION DENUDER; DRY DEPOSITION; AIR CONCENTRATION;
D O I
10.1016/0960-1686(93)90280-C
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A new diffusion denuder is described for the continuous measurement of atmospheric ammonia. Ammonia is collected in an absorption solution in a rotating denuder, separated from interfering compounds by diffusion through a semi-permeable membrane and detected by conductometry. The method is free from interferences by other atmospheric gases, with the exception of volatile amines. The detection limit is 6 ng m-3 for a 30-min integration time. This compact instrument is fully automated and suited for routine deployment in field studies. The precision is sufficiently high for micrometeorological studies of air-surface exchange of ammonia.
引用
收藏
页码:2085 / 2090
页数:6
相关论文
共 19 条