ELECTRODE CURRENT DETERMINATION FROM PROGRAMMABLE VOLTAGE SOURCES

被引:10
作者
LIDGEY, FJ
ZHU, QS
MCLEOD, CN
BRECKON, WR
机构
来源
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT | 1992年 / 13卷
关键词
D O I
10.1088/0143-0815/13/A/009
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The accuracy requirements for adaptive current electrical impedance tomography (EIT) Measurements exceed the capability of available current sources. A new architecture for an EIT system is described in which the electrode current is set indirectly from a voltage-drive structure. A numerically inverted admittance matrix, obtained from current measurements and driving voltages, has been used to achieve the desired current pattern from programmable voltage sources.