SIMULTANEOUS CONTINUOUS-FLOW ANALYSIS OF FREE AND TOTAL SULFUR-DIOXIDE IN WINE

被引:16
作者
FALCONE, F
MAXWELL, KC
机构
[1] Quality Assurance Laboratory, Liquor Control Board of Ontario, Toronto, Ontario
关键词
D O I
10.1021/jf00020a013
中图分类号
S [农业科学];
学科分类号
09 ;
摘要
A rapid, reliable method (60 samples/h) for the simultaneous determination of free and total sulfur dioxide (SO2) in wine has been developed employing a third-generation continuous flow analyzer, the TRAACS 800. Continuous flow analysis measured 75-100 % free SO2 compared to the Ripper method. Recoveries from sulfite-spiked wine samples ranged from 62 to 104 %, with an average precision of 4.5 %.
引用
收藏
页码:1355 / 1357
页数:3
相关论文
共 14 条
[1]  
BUECHSENSTEIN JW, 1978, AM J ENOL VITICULT, V29, P161
[2]   DETERMINATION OF ATMOSPHERIC SULFUR-DIOXIDE WITHOUT TETRACHLOROMERCURATE(II) AND THE MECHANISM OF THE SCHIFF REACTION [J].
DASGUPTA, PK ;
DECESARE, K ;
ULLREY, JC .
ANALYTICAL CHEMISTRY, 1980, 52 (12) :1912-1922
[3]  
DEVRIES JW, 1986, J ASSOC OFF ANA CHEM, V69, P827
[4]  
MAGUIERIRA A, 1987, ANALYST, V112, P1569
[5]  
MOORE LF, 1987, AM J ENOL VITICULT, V38, P28
[6]  
Ough C. S., 1988, METHODS ANAL MUSS WI, P222
[7]  
ROBINSON D, 1991, American Journal of Enology and Viticulture, V42, P81
[8]   FLOW INJECTION ANALYSES .1. NEW CONCEPT OF FAST CONTINUOUS-FLOW ANALYSIS [J].
RUZICKA, J ;
HANSEN, EH .
ANALYTICA CHIMICA ACTA, 1975, 78 (01) :145-157
[9]  
STONYS DB, 1987, J ASSOC OFF ANA CHEM, V70, P114
[10]  
SULLIVAN JJ, 1990, J ASSOC OFF ANA CHEM, V73, P35