GAS-DIFFUSION SEPARATION AND FLOW-INJECTION POTENTIOMETRY - A FRUITFUL ALLIANCE OF ANALYTICAL METHODS

被引:34
作者
FRENZEL, W
机构
[1] Institut für Technischen Umweltschutz, Fachgebiet Luftreinhaltung, Technische Universität Berlin, Berlin 12, D-1000
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1990年 / 336卷 / 01期
关键词
D O I
10.1007/BF00348508
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The similarities and differences of the operation principle of gas-sensing electrodes and potentiometric detection coupled to gas-diffusion separation in flow injection analysis are discussed with special emphasis on selectivity and sensitivity aspects. Several examples of application are presented highlighting the improvements in detectability obtained by gas-diffusion flow injection potentiometry. High sensitivity determination of ammonium is achieved through accumulation of ammonia released from the sample stream in the small recipient volume of the gas-diffusion unit. A method for almost specific determination of cyanide is presented making use of gas-diffusion separation of hydrogen cyanide and potentiometric detection with a selective AgI membrane electrode. The interference of sulfide is totally prevented by its oxidation in the donor line. If applied to potentiometric measurement following gas diffusion separation an intrinsically non-selective metallic silver wire electrode turns out to enable the selective detection of sulfide with high sensitivity and fast response. A new approach for diffusive sampling and on-line detection of gas-phase contaminants is exemplified by the determination of NOx. © 1990 Springer-Verlag.
引用
收藏
页码:21 / 28
页数:8
相关论文
共 53 条