DESIGN AND IMPLEMENTATION OF A FIBER-OPTIC-BASED RESIDUAL CHLORINE MONITOR

被引:9
作者
MOUAZIZ, Z [1 ]
BRIGGS, R [1 ]
HAMILTON, I [1 ]
GRATTAN, KTV [1 ]
机构
[1] SEVERN TRENT ATEC,W MIDLANDS B76 9B1,ENGLAND
关键词
D O I
10.1016/0925-4005(93)85284-H
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The design and implementation of a prototype optical residual chlorine monitor (RCM) for the purpose of chlorine concentration measurement in continuous flow is reported. A precision of 0.1 mg/l measured as OCl- is reported and the upper limit of the measurement range can be extended from 1 to 20 mg/l with a change in concentration of an added reagent. A background reference to allow for the presence of other interfering species is obtained by taking the ratio of the signals measured before and after chlorine is removed from the sample, with the addition of sodium sulphite. This compensates for any lamp and electronic-circuit drifts. The prototype has been tested over a wider range of waters and long periods of up to 24 h.
引用
收藏
页码:431 / 440
页数:10
相关论文
共 15 条
[1]   CONTINUOUS-FLOW DETERMINATION OF FREE CHLORINE IN WATER [J].
AOKI, T ;
MUNEMORI, M .
ANALYTICAL CHEMISTRY, 1983, 55 (02) :209-212
[2]  
BALT L, 1981, ANAL CHEM, V3, P1799
[3]  
BRIGGS R, 1990, ADV WAT POL, P39
[4]  
GRATTAN KTV, 1987, J OPT SENSORS, V2, P378
[5]  
JOHNSON JD, 1978, J AM WATER WORKS ASS, V70, P341
[6]   STUDIES WITH IMMOBILIZED CHEMICAL REAGENTS USING A FLOW-CELL FOR THE DEVELOPMENT OF CHEMICALLY SENSITIVE FIBRE-OPTIC DEVICES [J].
KIRKBRIGHT, GF ;
NARAYANASWAMY, R ;
WELTI, NA .
ANALYST, 1984, 109 (01) :15-17
[7]   DETERMINATION OF CHLORINE IN WATER BY LUMINOL CHEMI-LUMINESCENCE [J].
MARINO, DF ;
INGLE, JD .
ANALYTICAL CHEMISTRY, 1981, 53 (03) :455-458
[8]   DETERMINATION OF RESIDUAL CHLORINE IN WATER WITH COMPUTER AUTOMATION AND A RESIDUAL-CHLORINE ELECTRODE [J].
RIGDON, LP ;
MOODY, GJ ;
FRAZER, JW .
ANALYTICAL CHEMISTRY, 1978, 50 (03) :465-469
[9]  
SCHIEGGI AM, 1987, J OPT SENSORS, V2, P3
[10]  