PERFORMANCE EVALUATION OF COMPUTED-TOMOGRAPHY SCANNERS USING A PHANTOM MODEL

被引:12
作者
BELLON, EM [1 ]
MIRALDI, FD [1 ]
WIESEN, EJ [1 ]
机构
[1] CLEVELAND METROPOLITAN GEN HOSP,CLEVELAND,OH 44109
关键词
D O I
10.2214/ajr.132.3.345
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A phantom was designed to test computed tomography (CT) scanner performance relative to spatial resolution, contrast resolution, field uniformity, spatial linearity, artifact resistance, and radiation dose. Translate/rotate, stationary detector with rotating X-ray tube, and rotate-only scanners were tested under conditions that produced the best possible results for each machine within dose ranges used in clinical practice. Machines can be ranked according to results for each parameter tested. The range of performance exhibited on machines tested was unrelated to class of scanner. Further, machine dedication to a particular body region did not necessarily secure superior results for that region. Spatial resolution at low contrast levels (poor on most machines) seems to be the most sensitive test of performance.
引用
收藏
页码:345 / 352
页数:8
相关论文
共 9 条
  • [1] COMPUTED TOMOGRAPHIC SCANNING IN CHILDREN - COMPARISON OF RADIATION-DOSE AND RESOLVING POWER OF COMMERCIAL CT SCANNERS
    BRASCH, RC
    BOYD, DP
    GOODING, CA
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1978, 131 (01) : 95 - 101
  • [2] COHEN G, 1977, NOV ANN M RAD SOC N
  • [3] DEVELOPMENT OF A PHANTOM FOR EVALUATION AND ASSURANCE OF IMAGE QUALITY IN CT SCANNING
    GOODENOUGH, DJ
    WEAVER, KE
    DAVIS, DO
    [J]. OPTICAL ENGINEERING, 1977, 16 (01) : 52 - 65
  • [4] GRODSTEIN GW, 1957, NBS583
  • [5] JOHNS HE, 1971, PHYSICS RADIOLOGY
  • [6] COMPARATIVE MODULATION TRANSFER-FUNCTIONS OF EMI AND DELTA SCANNERS
    MACINTYRE, WJ
    ALFIDI, RJ
    HAAGA, J
    CHERNAK, E
    MEANY, TF
    [J]. RADIOLOGY, 1976, 120 (01) : 189 - 191
  • [7] PERFORMANCE EVALUATION AND QUALITY ASSURANCE OF COMPUTED TOMOGRAPHY SCANNERS, WITH ILLUSTRATIONS FROM EMI, ACTA, AND DELTA SCANNERS
    MCCULLOUGH, EC
    PAYNE, JT
    BAKER, HL
    HATTERY, RR
    SHEEDY, PF
    STEPHENS, DH
    GEDGAUDUS, E
    [J]. RADIOLOGY, 1976, 120 (01) : 173 - 188
  • [8] INACCURACY IN COMPUTED TOMOGRAPHY - ENERGY-DEPENDENCE OF CT VALUES
    ZATZ, LM
    ALVAREZ, RE
    [J]. RADIOLOGY, 1977, 124 (01) : 91 - 97
  • [9] 1977, PHANTOMS PERFORMANCE