HOW ACCURATE ARE HOME BLOOD-GLUCOSE METERS WITH SPECIAL RESPECT TO THE LOW GLYCEMIC RANGE

被引:14
作者
MOBERG, E
LUNDBLAD, S
LINS, PE
ADAMSON, U
机构
[1] Department of Medicine, Danderyd Hospital, Stockholm
关键词
BLOOD-GLUCOSE METERS; SELF-MONITORING OF BLOOD GLUCOSE; HYPOGLYCEMIA; DIABETES-MELLITUS;
D O I
10.1016/0168-8227(93)90119-P
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Comparisons were made between four blood-glucose meters (Diascan(R), Glucometer II(R), Reflolux II(R) and ExacTech(R)) and a reference method (YSI) for the full, clinical, blood-glucose range and for the subranges <4.4 mmol/l, 4.4-10.0 mmol/l and > 10.0 mmol/l, respectively. In the low-glucose range, the error-grid analysis was also applied. All the meters showed acceptable agreement with the reference method when the whole glucose range was considered, yielding r-values between 0.96 and 0.99. However, when the results were separated in the different subranges, the outcome was different, in that the Diascan(R) meter displayed systematically high glucose levels and the ExacTech(R) meter showed a great spread of the values within the low-glucose range, whereas the Glucometer II(R) and the Reflolux II(R) meter were less accurate within the high-glucose range. By applying the error-grid analysis, several errors of clinical importance within the low glycemic range were revealed, while other significant errors, which might lead to inadequate therapeutic decisions, were classified as 'clinically accurate'.
引用
收藏
页码:239 / 243
页数:5
相关论文
共 10 条
  • [1] EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE
    CLARKE, WL
    COX, D
    GONDERFREDERICK, LA
    CARTER, W
    POHL, SL
    [J]. DIABETES CARE, 1987, 10 (05) : 622 - 628
  • [2] EVALUATION OF A NEW SYSTEM FOR SELF BLOOD-GLUCOSE MONITORING
    CLARKE, WL
    BECKER, DJ
    COX, D
    SANTIAGO, JV
    WHITE, NH
    BETSCHART, J
    ECKENRODE, K
    LEVANDOSKI, LA
    PRUSINSKI, EA
    SIMINEIRO, LM
    SNYDER, AL
    TIDEMAN, AM
    YAEGER, T
    [J]. DIABETES RESEARCH AND CLINICAL PRACTICE, 1988, 4 (03) : 209 - 214
  • [3] ACCURACY OF PERCEIVING BLOOD-GLUCOSE IN IDDM
    COX, DJ
    CLARKE, WL
    GONDERFREDERICK, L
    POHL, S
    HOOVER, C
    SNYDER, A
    ZIMBELMAN, L
    CARTER, WR
    BOBBITT, S
    PENNEBAKER, J
    [J]. DIABETES CARE, 1985, 8 (06) : 529 - 536
  • [4] COMPARISON OF 5 GLUCOSE METERS FOR SELF-MONITORING OF BLOOD-GLUCOSE BY DIABETIC-PATIENTS
    GIFFORDJORGENSEN, RA
    BORCHERT, J
    HASSANEIN, R
    TILZER, L
    EAKS, GA
    MOORE, WV
    [J]. DIABETES CARE, 1986, 9 (01) : 70 - 76
  • [5] ARTERIAL, ARTERIALIZED VENOUS, VENOUS AND CAPILLARY BLOOD-GLUCOSE MEASUREMENTS IN NORMAL MAN DURING HYPERINSULINEMIC EUGLYCEMIA AND HYPOGLYCEMIA
    LIU, D
    MOBERG, E
    KOLLIND, M
    LINS, PE
    ADAMSON, U
    MACDONALD, IA
    [J]. DIABETOLOGIA, 1992, 35 (03) : 287 - 290
  • [6] HOME MONITORS OF BLOOD-GLUCOSE - COMPARISON OF PRECISION AND ACCURACY
    NORTH, DS
    STEINER, JF
    WOODHOUSE, KM
    MADDY, JA
    [J]. DIABETES CARE, 1987, 10 (03) : 360 - 366
  • [7] SHORT-TERM EVALUATION OF AN ELECTROCHEMICAL SYSTEM (EXACTECHTM) FOR BLOOD-GLUCOSE MONITORING
    ROSS, D
    HEINEMANN, L
    CHANTELAU, EA
    [J]. DIABETES RESEARCH AND CLINICAL PRACTICE, 1990, 10 (03) : 281 - 285
  • [8] ACCURACY OF HOME BLOOD-GLUCOSE MONITORS
    TATE, PF
    CLEMENTS, CA
    WALTERS, JE
    [J]. DIABETES CARE, 1992, 15 (04) : 536 - 538
  • [9] FREQUENCY OF DAYTIME BIOCHEMICAL HYPOGLYCEMIA IN INSULIN-TREATED DIABETIC-PATIENTS - RELATION TO DAILY MEDIAN BLOOD-GLUCOSE CONCENTRATIONS
    THORSTEINSSON, B
    PRAMMING, S
    LAURITZEN, T
    BINDER, C
    [J]. DIABETIC MEDICINE, 1986, 3 (02) : 147 - 151
  • [10] 1987, IDF B, V32, P169