MATHEMATICAL INDEX OF UNIFORMITY (IOU) FOR SENSITIVITY AND RESOLUTION

被引:4
作者
NUSYNOWITZ, ML [1 ]
BENEDETTO, AR [1 ]
机构
[1] WILLIAM BEAUMONT ARMY MED CTR,CLIN INVEST SERV,EL PASO,TX 79920
关键词
D O I
10.1148/131.1.235
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors define a mathematically precise index of uniformity (IOU) and show that the IOU is a convenient tool to quantitate uniformity of sensitivity and resolution response for any given area of interest on the detector face. They also show that the IOU is an appropriate factor for weighing the mean values of sensitivity and resolution for nonuniformity of response; these weighted sensitivity and resolution factors are used to derive a weighted figure of merit which takes into account, in a single parameter, the nonuniformities of response inherent in the determination of sensitivity and resolution.
引用
收藏
页码:235 / 241
页数:7
相关论文
共 18 条
[1]   QUANTITATIVE ASSESSMENT OF FIELD UNIFORMITY FOR GAMMA CAMERAS [J].
COHEN, G ;
KEREIAKES, JG ;
PADIKAL, TN ;
ASHARE, AB ;
SAENGER, EL .
RADIOLOGY, 1976, 118 (01) :197-200
[2]  
CRADDUCK TD, 1977, NUCLEAR MEDICINE PHY, P456
[3]  
CROXTON FE, 1953, ELEMENTARY STATISTIC, P159
[4]  
GOTTSCHALK A, 1968, FUNDAMENTAL PROBL ED, P301
[5]   PITFALLS IN GAMMA CAMERA FIELD UNIFORMITY CORRECTION [J].
HANNAN, WJ ;
BESSENT, RG .
BRITISH JOURNAL OF RADIOLOGY, 1974, 47 (563) :820-820
[6]   PITFALLS IN GAMMA CAMERA FIELD UNIFORMITY CORRECTION [J].
JANSSON, LG ;
PARKER, RP .
BRITISH JOURNAL OF RADIOLOGY, 1975, 48 (569) :408-409
[7]  
KEYES JW, 1972, J NUCL MED, V13, P525
[8]  
KIRCH DL, 1978, J NUCL MED, V19, P712
[9]  
KREYSZIG E, 1967, ADV ENGINEERING MATH, P384
[10]  
MORRISON LM, 1971, J NUCL MED, V12, P785