RELIABILITY OF THE DINAMAP NONINVASIVE MONITOR IN THE MEASUREMENT OF BLOOD-PRESSURE OF III ASIAN NEWBORNS

被引:22
作者
CHIA, F
ANG, AT
WONG, TW
TAN, KW
FUNG, KP
LEE, J
KHIN, K
机构
[1] CHINESE UNIV HONG KONG,DEPT COMMUNITY MED,SHA TIN,HONG KONG
[2] NATL UNIV SINGAPORE,DEPT PEDIAT,SINGAPORE,SINGAPORE
[3] NATL UNIV SINGAPORE,DEPT COMMUNITY OCCUPAT & FAMILY MED,SINGAPORE 0511,SINGAPORE
关键词
D O I
10.1177/000992289002900502
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Four hundred thirty-one paired sets of readings of systolic and diastolic blood pressure and 438 paired sets of readings of mean arterial BP from 49 ill newborns, including 21 very low birth weight infants, were analyzed for the extent and pattern of agreement and the linear relationship between the Dinamap oscillometric monitor and the direct intraarterial blood pressure readings. Agreement between the two methods was measured by the intraclass correlation, whereas the linear relationship was assessed by the product-moment correlation. The intraclass correlations for systolic, diastolic and mean blood pressures were 0.696, 0.766, and 0.781, respectively. The product-moment correlations for systolic, diastolic and mean blood pressures were 0.706, 0.768, and 0.786, respectively. BP measurements by the Dinamap monitor showed reasonably close agreement to those obtained by the intraarterial mean arterial pressure ranges above 40 mmHg. For mean arterial pressure of 40 mmHg and lower, BP readings by the Dinamap monitor tended to be higher than those obtained by the intraarterial method. These findings appeared to be consistent regardless of the birth weight of the newborn. © 1990, Sage Publications. All rights reserved.
引用
收藏
页码:262 / 267
页数:6
相关论文
共 13 条
[1]   BLOOD-PRESSURE MONITORING IN NEONATES - COMPARISON OF UMBILICAL AND PERIPHERAL ARTERY CATHETER MEASUREMENTS [J].
BUTT, WW ;
WHYTE, H .
JOURNAL OF PEDIATRICS, 1984, 105 (04) :630-632
[2]  
DELLAGRAMMATICAS H D, 1981, Clinical Physics and Physiological Measurement, V2, P271, DOI 10.1088/0143-0815/2/4/003
[3]   DINAMAP FAILS TO DETECT HYPOTENSION IN VERY-LOW-BIRTH-WEIGHT INFANTS [J].
DIPROSE, GK ;
EVANS, DH ;
ARCHER, LNJ ;
LEVENE, MI .
ARCHIVES OF DISEASE IN CHILDHOOD, 1986, 61 (08) :771-773
[4]  
FUNG KP, IN PRESS COMP APPL B
[5]   AORTIC BLOOD PRESSURE IN INFANTS ADMITTED TO A NEONATAL INTENSIVE CARE UNIT [J].
HALL, RT .
AMERICAN JOURNAL OF DISEASES OF CHILDREN, 1971, 121 (02) :145-&
[6]   AN AUTOMATED OSCILLOMETRIC TECHNIQUE FOR ESTIMATING MEAN ARTERIAL-PRESSURE IN CRITICALLY ILL NEWBORNS [J].
KIMBLE, KJ ;
DARNALL, RA ;
YELDERMAN, M ;
ARIAGNO, RL ;
REAM, AK .
ANESTHESIOLOGY, 1981, 54 (05) :423-425
[8]   STATISTICAL EVALUATION OF AGREEMENT BETWEEN 2 METHODS FOR MEASURING A QUANTITATIVE VARIABLE [J].
LEE, J ;
KOH, D ;
ONG, CN .
COMPUTERS IN BIOLOGY AND MEDICINE, 1989, 19 (01) :61-70
[9]   NUTRIENT INTAKES OF HUSBANDS AND WIVES - IMPLICATIONS FOR EPIDEMIOLOGIC RESEARCH [J].
LEE, J ;
KOLONEL, LN .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 1982, 115 (04) :515-525
[10]  
LUI K, 1982, AUST PAEDIATR J, V18, P32