THE SAFETY AND EFFICACY OF CHRONIC VENTRICULAR PACING AT 1.6 VOLTS USING A STEROID ELUTING LEAD

被引:26
作者
STAMATO, NJ
OTOOLE, MF
FETTER, JG
ENGER, EL
机构
[1] MIDWEST CARDIOVASC INST,LOMBARD,IL
[2] MEDTRONIC INC,MINNEAPOLIS,MN 55418
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1992年 / 15卷 / 03期
关键词
PACEMAKER; STEROID LEADS;
D O I
10.1111/j.1540-8159.1992.tb06491.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Steroid eluting leads may allow for lower chronic pacing thresholds and therefore lower pacing outputs. Twenty-two patients (15 presenting with syncope) were implanted with VVI or VVIR pacemakers and transvenous steroid eluting leads and followed for a mean of 20.6 months while being paced at 1.6 V and 0.6 msec. Mean acute voltage pacing thresholds were 0.40 V at 0.5 msec and chronic pulse width thresholds were 0.21 msec at 0.8 V. Pacemaker function was documented with one to three 24-hour Holter monitors, attached during the 2-6 week postimplant period, bimonthly transtelephonic monitoring, and monthly pacemaker clinic visits. No patient developed recurrent symptoms and consistent capture was verified in all patients on every 24-hour Holter recording and transtelephonic monitor. Chronic ventricular pacing at an output of 1.6 V at 0.6 msec is safe and effective when using a steroid eluting lead and potentially has implications for pacemaker longevity.
引用
收藏
页码:248 / 251
页数:4
相关论文
共 13 条
[1]  
ALAREZNOVOA J, 1985, CARDIAC PACING, P436
[2]  
ASTRINSKY EA, 1983, CLIN PROG PACING ELE, V1, P51
[3]   CELLULAR ELECTROPHYSIOLOGIC EFFECTS OF DEXAMETHASONE SODIUM-PHOSPHATE - IMPLICATIONS FOR CARDIAC STIMULATION WITH STEROID-ELUTING ELECTRODES [J].
BENDITT, DG ;
KRIETT, JM ;
RYBERG, C ;
GORNICK, CC ;
STOKES, KL ;
DETLOFF, BLS .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1989, 22 (01) :67-73
[4]   SENSOR-TRIGGERED, RATE-VARIABLE CARDIAC PACING - CURRENT TECHNOLOGIES AND CLINICAL IMPLICATIONS [J].
BENDITT, DG ;
MILSTEIN, S ;
BUETIKOFER, J ;
GORNICK, CC ;
MIANULLI, M ;
FETTER, J .
ANNALS OF INTERNAL MEDICINE, 1987, 107 (05) :714-724
[5]  
FETTER J, 1990, PACE, V13, P525
[6]   PACEMAKER LONGEVITY [J].
FURMAN, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (09) :1437-1438
[7]  
Hoff P.I., 1985, CARDIAC PACING ELECT, P1014
[8]   THE POROUS TITANIUM STEROID ELUTING ELECTRODE - A DOUBLE-BLIND STUDY ASSESSING THE STIMULATION THRESHOLD EFFECTS OF STEROID [J].
MOND, H ;
STOKES, K ;
HELLAND, J ;
GRIGG, L ;
KERTES, P ;
PATE, B ;
HUNT, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (02) :214-219
[9]  
SCHUCHERT A, 1989, PACE, V12, P680
[10]  
SCHUCHERT A, 1989, EUR HEART J, V19, P139