INITIAL CLINICAL-EXPERIENCE IN MR-IMAGING OF THE BRAIN WITH A FAST FLUID-ATTENUATED INVERSION-RECOVERY PULSE SEQUENCE

被引:243
作者
RYDBERG, JN
HAMMOND, CA
GRIMM, RC
ERICKSON, BJ
JACK, CR
HUSTON, J
RIEDERER, SJ
机构
[1] MAYO CLIN & MAYO FDN,DEPT DIAGNOST RADIOL,MAGNET RESONANCE LAB,ROCHESTER,MN 55905
[2] UNIV MINNESOTA,DEPT BIOMED ENGN,MINNEAPOLIS,MN 55455
关键词
BRAIN; ABNORMALITIES; MR; MAGNETIC RESONANCE (MR); PULSE SEQUENCES;
D O I
10.1148/radiology.193.1.8090888
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE To evaluate fast fluid-attenuated inversion-recovery (FLAIR) technique for imaging brain abnormalities. MATERIALS AND METHODS: A fast FLAIR sequence was developed that provided 36 5-mm contiguous sections in 5 minutes 8 seconds. Resulting images were compared with dual-echo T2-weighted spin-echo images of 41 consecutive patients with brain abnormalities. RESULTS: Contrast and contrast-to-noise ratios (C/Ns) (for contrast between the lesion and background and between the lesion and cerebrospinal fluid) for fast FLAIR exceeded the corresponding values for T2-weighted spin-echo images for all but the second-echo lesion-to-background C/N. Fast FLAIR provided equivalent or greater overall lesion conspicuity and enabled greater lesion detection in 98% and 100%, respectively, of the evaluations. Fast FLAIR images more often had image artifact, but this did not interfere with image interpretation in a significantly (P less than or equal to .05) greater number of evaluations. CONCLUSION: Fast FLAIR provides images that are superior to proton-density- and T2-weighted images for many image quality criteria.
引用
收藏
页码:173 / 180
页数:8
相关论文
共 28 条
[1]  
BOTTOMLEY PA, 1984, MED PHYS, V11, P425, DOI 10.1118/1.595535
[2]   MAGNETIC-RESONANCE OF THE BRAIN - THE OPTIMAL SCREENING TECHNIQUE [J].
BRANTZAWADZKI, M ;
NORMAN, D ;
NEWTON, TH ;
KELLY, WM ;
KJOS, B ;
MILLS, CM ;
DILLON, W ;
SOBEL, D ;
CROOKS, LE .
RADIOLOGY, 1984, 152 (01) :71-77
[3]   T2-WEIGHTED SPIN-ECHO PULSE SEQUENCE WITH VARIABLE REPETITION AND ECHO TIMES FOR REDUCTION OF MR IMAGE ACQUISITION TIME [J].
BUTTS, RK ;
FARZANEH, F ;
RIEDERER, SJ ;
RYDBERG, JN ;
GRIMM, RC .
RADIOLOGY, 1991, 180 (02) :551-556
[4]  
CASPERS JM, 1993, RADIOLOGY, V189, P391
[5]   THE LOSS OF SMALL OBJECTS IN VARIABLE TE IMAGING - IMPLICATIONS FOR FSE, RARE, AND EPI [J].
CONSTABLE, RT ;
GORE, JC .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) :9-24
[6]   MRI OF THE BRAIN-STEM USING FLUID ATTENUATED INVERSION RECIVERY PULSE SEQUENCES [J].
DECOENE, B ;
HAJNAL, JV ;
PENNOCK, JM ;
BYDDER, GM .
NEURORADIOLOGY, 1993, 35 (05) :327-331
[7]  
DECOENE B, 1992, AM J NEURORADIOL, V13, P1555
[8]  
DENBOER JA, 1993, SOC MAGNETIC RESONAN, P328
[9]   HIGH SIGNAL REGIONS IN NORMAL WHITE MATTER SHOWN BY HEAVILY T2-WEIGHTED CSF NULLED IR SEQUENCES [J].
HAJNAL, JV ;
DECOENE, B ;
LEWIS, PD ;
BAUDOUIN, CJ ;
COWAN, FM ;
PENNOCK, JM ;
YOUNG, IR ;
BYDDER, GM .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (04) :506-513
[10]  
HAJNAL JV, 1992, SOC MAGNETIC RESONAN, P1007