BEAM CHARACTERISTICS OF THE THERAPAX DXT300 ORTHOVOLTAGE THERAPY UNIT

被引:29
作者
GERIG, L
SOUBRA, M
SALHANI, D
机构
[1] CARLETON UNIV,OTTAWA K1S 5B6,ON,CANADA
[2] UNIV OTTAWA,OTTAWA,ON,CANADA
关键词
D O I
10.1088/0031-9155/39/9/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The operating performance and beam characteristics of a new orthovoltage unit, the Therapax DXT300, have been evaluated. Percentage depth-dose and backscatter tables are presented for several applicator sizes, at 30 cm and 50 cm focal skin distances (FSDS) and for multiple x-ray beam qualities with the tube operating between 100 and 300 kVp accelerating potential. The unit has been found to provide beam characteristics similar to those reported for other orthovoltage therapy machines. The linearity and short- and long-term stability/reproducibility of the unit's internal dosimetry system have also been studied, and results indicate a very stable beam output of better than 1% standard deviation. The data presented in this work should provide the basis for comparison with other units and act as a reference for clinics commissioning the Therapax DXT300 in the future.
引用
收藏
页码:1377 / 1392
页数:16
相关论文
共 12 条
[1]   DIFFERENCES IN REPORTED BACKSCATTER FACTORS FOR LOW-ENERGY X-RAYS - A LITERATURE STUDY [J].
CARLSSON, CA .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (04) :521-531
[2]   DEPENDENCE OF THE PHOTON BACKSCATTER FACTOR FOR WATER ON SOURCE-TO-PHANTOM DISTANCE AND IRRADIATION FIELD SIZE [J].
GROSSWENDT, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1233-1245
[3]   DEPENDENCE OF THE PHOTON BACKSCATTER FACTOR FOR WATER ON IRRADIATION FIELD SIZE AND SOURCE-TO-PHANTOM DISTANCES BETWEEN 1.5 AND 10 CM [J].
GROSSWENDT, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (02) :305-310
[5]  
KURUP RG, 1993, MED DOSIM, V8, P179
[6]  
NIMROOMANDRAD A, 1987, MED PHYS, V14, P874
[7]   PERFORMANCE-CHARACTERISTICS OF A WIDELY USED ORTHOVOLTAGE X-RAY THERAPY MACHINE [J].
SCRIMGER, JW ;
CONNORS, SG .
MEDICAL PHYSICS, 1986, 13 (02) :267-269
[8]  
TSIEN KC, 1962, ISODOSE CHARTS TABLE
[9]  
1983, BR J RADIOL S17
[10]  
1991, PHYS MED BIOL, V36, P1027