DESIGN AND PROPERTIES OF A FLOW-INJECTION ANALYSIS CELL USING POTASSIUM-SELECTIVE ION-SENSITIVE FIELD-EFFECT TRANSISTORS AS DETECTION ELEMENTS

被引:27
作者
VANDERWAL, PD
SUDHOLTER, EJR
REINHOUDT, DN
机构
[1] Laboratory of Organic Chemistry, University of Twente, 7500 AE Enschede
关键词
FLOW SYSTEM; VOLTAMMETRY; ION-SENSITIVE FIELD-EFFECT TRANSISTORS; POTASSIUM; SERUM; URINE;
D O I
10.1016/S0003-2670(00)80216-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The combination of flow-injection analysis (FIA) and chemically modified ion-sensitive field-effect transistors (CHEMFETs) is described. In a wall-jet cell, two identical potassium-selective CHEMFETs were used for a differential measurement using a platinum (pseudo-)reference electrode. Silicone-rubber membrane materials, chemically bound to the SiO2 gate oxide, were used with valinomycin as the ionophore. The optimized FIA system showed a linear response of 56 mV per decade for potassium concentrations above 5 x 10(-5) M. Preliminary results of potassium determinations in human serum and urine samples are presented.
引用
收藏
页码:159 / 166
页数:8