ANTERIOR CHEEK ELECTRODES ARE COMPARABLE TO SPHENOIDAL ELECTRODES FOR THE IDENTIFICATION OF ICTAL ACTIVITY

被引:33
作者
KRAUSS, GL [1 ]
LESSER, RP [1 ]
FISHER, RS [1 ]
ARROYO, S [1 ]
机构
[1] JOHNS HOPKINS UNIV HOSP, DEPT NEUROL, BALTIMORE, MD 21205 USA
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1992年 / 83卷 / 06期
关键词
EPILEPSY; ELECTROENCEPHALOGRAM; ICTAL RECORDING; SPHENOIDAL ELECTRODES;
D O I
10.1016/0013-4694(92)90068-S
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Sphenoidal electrodes are used to localize epileptiform activity originating in the temporal lobe during complex partial seizures. Sphenoidal electrodes, however, are semi-invasive and uncomfortable to the patient. We compared skin electrodes placed on the cheek (''cheek electrodes'') with sphenoidal electrodes for the detection of the side and site of complex partial seizure onset. In a masked, randomized comparison of single ictal recordings in 22 patients, there were no significant differences between sphenoidal and cheek electrode montages in detecting the side or site of ictal onset (P < 0.01 ). Signal/noise ratios for interictal spikes were a mean 16.5% greater at sphenoidal sites compared to cheek sites (paired t test, t = 2.4, P < 0.05). This difference, however, did not influence the detection of rhythmical ictal activity in cheek and sphenoidal montages in our study, nor the assignment of side, site or time of seizure onset by unbiased readers. Recordings from cheek electrodes are comparable to those from sphenoidal electrodes and are useful for localizing ictal activity.
引用
收藏
页码:333 / 338
页数:6
相关论文
共 22 条
[1]   DISTRIBUTION OF TEMPORAL SPIKES IN RELATION TO THE SPHENOIDAL ELECTRODE [J].
BINNIE, CD ;
MARSTON, D ;
POLKEY, CE ;
AMIN, D .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1989, 73 (05) :403-409
[2]  
BLUME WT, 1984, ELECTROEN CLIN NEURO, V57, P295, DOI 10.1016/0013-4694(84)90151-2
[3]   STANDARD SPHENOIDAL AND MINI-SPHENOIDAL ELECTRODES - A DIRECT COMPARISON [J].
BUCHHALTER, JR ;
SCHOMER, DL ;
SCHACHTER, SC .
JOURNAL OF EPILEPSY, 1991, 4 (01) :29-31
[4]  
ENGEL J, 1989, SEIZURES EPILEPSY, P150
[5]   COMPARISON OF TRADITIONAL READING OF EEG AND AUTOMATIC RECOGNITION OF INTERICTAL EPILEPTIC ACTIVITY [J].
GOTMAN, J ;
GLOOR, P ;
SCHAUL, N .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1978, 44 (01) :48-60
[6]   ANTERIOR TEMPORAL ELECTRODES IN COMPLEX PARTIAL SEIZURES [J].
HOMAN, RW ;
JONES, MC ;
RAWAT, S .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1988, 70 (02) :105-109
[7]   CEREBRAL LOCATION OF INTERNATIONAL 10-20 SYSTEM ELECTRODE PLACEMENT [J].
HOMAN, RW ;
HERMAN, J ;
PURDY, P .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1987, 66 (04) :376-382
[8]   NEW SPHENOIDAL ELECTRODE ASSEMBLY TO PERMIT LONG-TERM MONITORING OF PATIENTS ICTAL OR INTERICTAL EEG [J].
IVES, JR ;
GLOOR, P .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1977, 42 (04) :575-580
[9]  
IVES JR, 1990, EPILEPSIA, V31, P662
[10]  
JONES DP, 1951, ELECTROEN CLIN NEURO, V3, P100