DETERMINATION OF CYCLOSPORINE A IN WHOLE-BLOOD BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING AUTOMATED COLUMN SWITCHING

被引:7
作者
HOSOTSUBO, H [1 ]
TAKEZAWA, J [1 ]
TAENAKA, N [1 ]
HOSOTSUBO, K [1 ]
YOSHIYA, I [1 ]
机构
[1] OSAKA UNIV HOSP, INTENS CARE UNIT, FUKUSHIMA KU, OSAKA, JAPAN
来源
JOURNAL OF CHROMATOGRAPHY | 1986年 / 383卷 / 02期
关键词
D O I
10.1016/S0378-4347(00)83480-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fully automated high-performance liquid chromatographic column-switching system is presented for the determination of cyclosporin A in whole blood. After blood proteins were precipitated with acetonitrile, the supernatant was automatically loaded on to a cyanopropyl column for initial separation, and then the fraction containing cyclosporin A was loaded on the a trimethylsilica column for final separation and quantitation. Cyclosporin A was detected by ultraviolet absorption at 205 nm. The minimum detectable concentration of cyclosporin A was 5 ng/ml in 100 .mu.l of blood. The coefficient of variation of the method was 1.755, 1.748 and 0.655% in whole blood when spiked at the 170, 425 and 850 ng/ml levels, respectively. One assay was completed in 15 min.
引用
收藏
页码:349 / 355
页数:7
相关论文
共 17 条
[1]  
BANDINI G, 1983, LANCET, V1, P762
[2]  
BENNETT MJ, 1984, CLIN CHEM, V30, P817
[3]  
CARRUTHERS SG, 1983, CLIN CHEM, V29, P180
[4]   ANALYSIS OF THEOPHYLLINE BY AUTOMATED MULTIDIMENSIONAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY INVOLVING DIRECT PLASMA INJECTION [J].
CHANG, SF ;
WELSCHER, TM ;
OBER, RE .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1983, 72 (03) :236-239
[5]  
Donatsch P, 1981, J Immunoassay, V2, P19, DOI 10.1080/01971528108062989
[6]   SIMPLIFIED LIQUID-CHROMATOGRAPHIC ANALYSIS FOR CYCLOSPORIN-A IN BLOOD AND PLASMA WITH USE OF RAPID EXTRACTION [J].
GARRAFFO, R ;
LAPALUS, P .
JOURNAL OF CHROMATOGRAPHY, 1985, 337 (02) :416-422
[7]   MODIFIED COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE MEASUREMENT OF CYCLOSPORINE IN SERUM [J].
GMUR, DJ ;
YEE, GC ;
KENNEDY, MS .
JOURNAL OF CHROMATOGRAPHY, 1985, 344 (NOV) :422-427
[8]   SEMI-AUTOMATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF CYCLOSPORINE A IN WHOLE-BLOOD USING ONE-STEP SAMPLE PURIFICATION AND COLUMN-SWITCHING [J].
HAMILTON, G ;
ROTH, E ;
WALLISCH, E ;
TICHY, F .
JOURNAL OF CHROMATOGRAPHY, 1985, 341 (02) :411-419
[9]  
KABRA PM, 1985, CLIN CHEM, V31, P1717
[10]   SIMPLE AND RAPID HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF CYCLOSPORINE IN HUMAN-BLOOD AND SERUM [J].
KATES, RE ;
LATINI, R .
JOURNAL OF CHROMATOGRAPHY, 1984, 309 (02) :441-447