CONTINUOUS MEASUREMENT OF CARDIAC-OUTPUT DURING AORTIC CROSS-CLAMPING BY THE ESOPHAGEAL DOPPLER MONITOR ODM-1

被引:37
作者
KLOTZ, KF
KLINGSIEK, S
SINGER, M
WENK, H
ELEFTHERIADIS, S
KUPPE, H
SCHMUCKER, P
机构
[1] UNIV LUBECK,DEPT SURG,D-23538 LUBECK,GERMANY
[2] UNIV LONDON UNIV COLL,BLOOMSBURY INST INTENS CARE MED,SCH MED,LONDON WC1E 6JJ,ENGLAND
关键词
MEASUREMENT TECHNIQUES; CARDIAC OUTPUT; DOPPLER ECHOCARDIOGRAPHY;
D O I
10.1093/bja/74.6.655
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have compared the Doppler against the thermodilution technique far measurement of cardiac output in six patients during aortic surgery. The correlation coefficient between the two methods was between 0.76 and 0.84 during the different periods of the operation. Using the integral nomogram instead of direct calibration, the Doppler system underestimated cardiac output in atherosclerotic patients. However, the Doppler consent. Patients were excluded if they had known method did register accurately significant changes cardiac valve defects or any history of oesophageal in cardiac output.
引用
收藏
页码:655 / 660
页数:6
相关论文
共 29 条
  • [1] MYOCARDIAL ISCHEMIA DUE TO INFRARENAL AORTIC CROSS-CLAMPING DURING AORTIC SURGERY IN PATIENTS WITH SEVERE CORONARY-ARTERY DISEASE
    ATTIA, RR
    MURPHY, JD
    SNIDER, M
    LAPPAS, DG
    DARLING, RC
    LOWENSTEIN, E
    [J]. CIRCULATION, 1976, 53 (06) : 961 - 965
  • [2] NONINVASIVE CARDIAC-OUTPUT, DOPPLER FLOWMETRY, AND GOLD-PLATED ASSUMPTIONS
    BERNSTEIN, DP
    [J]. CRITICAL CARE MEDICINE, 1987, 15 (09) : 886 - 888
  • [3] CONTINUOUS NONINVASIVE REAL-TIME MONITORING OF STROKE VOLUME AND CARDIAC-OUTPUT BY THORACIC ELECTRICAL BIOIMPEDANCE
    BERNSTEIN, DP
    [J]. CRITICAL CARE MEDICINE, 1986, 14 (10) : 898 - 901
  • [4] BUSH HL, 1977, ARCH SURG-CHICAGO, V112, P1301
  • [5] SIMULTANEOUS MEASUREMENT OF CARDIAC-OUTPUT BY THERMODILUTION, THORACIC ELECTRICAL BIOIMPEDANCE AND DOPPLER ULTRASOUND
    CASTOR, G
    KLOCKE, RK
    STOLL, M
    HELMS, J
    NIEDERMARK, I
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1994, 72 (01) : 133 - 138
  • [6] SAFETY OF TRANSESOPHAGEAL ECHOCARDIOGRAPHY - A MULTICENTER SURVEY OF 10,419 EXAMINATIONS
    DANIEL, WG
    ERBEL, R
    KASPER, W
    VISSER, CA
    ENGBERDING, R
    SUTHERLAND, GR
    GRUBE, E
    HANRATH, P
    MAISCH, B
    DENNIG, K
    SCHARTL, M
    KREMER, P
    ANGERMANN, C
    ILICETO, S
    CURTIUS, JM
    MUGGE, A
    [J]. CIRCULATION, 1991, 83 (03) : 817 - 821
  • [7] COMPARISON OF PULSED DOPPLER AND THERMODILUTION METHODS FOR MEASURING CARDIAC-OUTPUT IN CRITICALLY ILL PATIENTS
    DONOVAN, KD
    DOBB, GJ
    NEWMAN, MA
    HOCKINGS, BES
    IRELAND, M
    [J]. CRITICAL CARE MEDICINE, 1987, 15 (09) : 853 - 857
  • [8] HEMODYNAMIC EVIDENCE FOR CARDIAC STRESS DURING TRANSURETHRAL PROSTATECTOMY
    EVANS, JWH
    SINGER, M
    CHAPPLE, CR
    MACARTNEY, N
    WALKER, JM
    MILROY, EJG
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 1992, 304 (6828): : 666 - 671
  • [9] BLOOD-VOLUME REDISTRIBUTION DURING CROSS-CLAMPING OF THE DESCENDING AORTA
    GELMAN, S
    KHAZAELI, MB
    ORR, R
    HENDERSON, T
    [J]. ANESTHESIA AND ANALGESIA, 1994, 78 (02) : 219 - 224
  • [10] GELMAN S, 1983, J THORAC CARDIOV SUR, V85, P287