DOSIMETRIC PROBLEMS AT LOW MONITOR UNIT SETTINGS FOR SCANNED AND SCATTERING FOIL ELECTRON-BEAMS

被引:13
作者
DAS, IJ
HARRINGTON, JC
AKBER, SF
TOMER, AF
MURRAY, JC
CHENG, CW
机构
[1] RADIAT ONCOL SERV RHODE ISL,WARWICK,RI 02886
[2] MARSHFIELD CLIN FDN MED RES & EDUC,DEPT RADIOL,MARSHFIELD,WI 54449
[3] UNIV MASSACHUSETTS,MED CTR,DEPT RADIAT ONCOL,WORCESTER,MA 01655
[4] UNIV ARIZONA,ARIZONA HLTH SCI CTR,DEPT RADIAT ONCOL,TUCSON,AZ 85724
关键词
MONITOR UNIT; ELECTRON BEAM; SCATTERING FOIL; SCANNED BEAM; DOSE LINEARITY; END EFFECT;
D O I
10.1118/1.597328
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Electron beam dosimetry at low monitor unit (MU) settings is important for dosimetric applications. Dose linearity, beam flatness, and beam energies were studied at low MU settings with various dose rates for different types of linear accelerators. It is observed that for the scattering foil units, the dose/MU is a smooth function of MU for all beam energies. Discrepancies in dose/MU are highest at the lowest MU. Significant variation (5%-245%) in dose linearity is observed among various linear accelerators at low MU settings. Dose rate has no effect on the dose linearity for all energies for the scattering foil units tested. On the contrary, for the scanned beam, there is no predictable pattern as dose/MU is random in nature and varies with time and beam energy. The maximum dosimetric error is observed for the highest energy beam where the beam width is most narrow. Using film, the beam uniformity was noticed to be very poor at low MU and high energy for scanned beams. The beam uniformity and dose linearity are random at low MU due to the random nature of the scan cycle. Under the adverse conditions, the deviation in dosimetric parameters was observed up to 200 MU.
引用
收藏
页码:821 / 826
页数:6
相关论文
共 17 条
[2]  
[Anonymous], 1983, MED PHYS, V10, P741
[3]   TELETHERAPY BEAM CHARACTERISTICS - THE 1ST 2ND [J].
BARISH, RJ ;
FLEISCHMAN, RC ;
PIPMAN, YM .
MEDICAL PHYSICS, 1987, 14 (04) :657-661
[5]   RECOMBINATION CORRECTION FACTORS FOR AN IONIZATION-CHAMBER EXPOSED TO DISCRETE PATTERNED PULSED SWEPT BEAMS [J].
CHANG, KS .
MEDICAL PHYSICS, 1993, 20 (02) :337-340
[6]  
CHEN E, 1993, MED PHYS, V20, P888
[7]   DOSIMETRIC ACCURACY AT LOW MONITOR UNIT SETTINGS [J].
DAS, IJ ;
KASE, KR ;
TELLO, VM .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (765) :808-811
[8]   ENERGY DETERMINATION FOR ELECTRON-BEAM CALIBRATION [J].
FREIM, JO ;
FELDMAN, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (03) :361-365
[9]  
JENSEN JM, 1990, STRAHLENTHER ONKOL, V166, P415
[10]   CLINICAL ELECTRON-BEAM DOSIMETRY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO-25 [J].
KHAN, FM ;
DOPPKE, KP ;
HOGSTROM, KR ;
KUTCHER, GJ ;
NATH, R ;
PRASAD, SC ;
PURDY, JA ;
ROZENFELD, M ;
WERNER, BL .
MEDICAL PHYSICS, 1991, 18 (01) :73-109