TREATMENT OF THE UNSTABLE INTERTROCHANTERIC FRACTURE - EFFECT OF THE PLACEMENT OF THE SCREW, ITS ANGLE OF INSERTION, AND OSTEOTOMY

被引:58
作者
DENHARTOG, BD
BARTAL, E
COOKE, F
机构
[1] UNIV KANSAS,SCH MED,DEPT SURG,ORTHOPAED & REHABIL SECT,WICHITA,KS
[2] ST FRANCIS & AFFILIATED HOSP,ORTHOPAED RESIDENCY PROGRAM,WICHITA,KS
关键词
D O I
10.2106/00004623-199173050-00011
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We created unstable intertrochanteric fractures with a large posteromedial defect in eighteen pairs of fresh-frozen femora from cadavera, and used the femora to evaluate the stability of six combinations of treatment with sliding compression screws and side-plates. The variables that were tested were high (150-degree) angle plates compared with low (130-degree) angle plates, the position of the lag-screw in the femoral head and neck (in the center of the head as seen on both anteroposterior and lateral roentgenograms compared with posteroinferiorly), and whether or not medial cortical contact had been re-established with a limited osteotomy of the greater trochanter. The femora were loaded to the point of failure. The mode of failure, maximum load to failure, and bending rigidity of each method of fixation were measured. Of the six types of treatment, the use of a 150-degree-angle plate, position of the lag-screw in the center of the head as seen on both anteroposterior and lateral roentgenograms, and osteotomy of the greater trochanter resulted in the highest mean load to failure and the greatest rigidity. Over-all, re-establishment of medial cortical contact by means of an osteotomy of the greater trochanter significantly improved the mean load to failure and the rigidity of the fixation. The higher-angle plate, although more difficult to insert, increased the mean load to failure but had no effect on rigidity. The position of the lag-screw (in the center of the head as seen on both anteroposterior and lateral roentgenograms or posteroinferiorly) had no bearing on the mean load to failure, but it significantly affected the mode of failure. CLINICAL RELEVANCE: The evidence suggested that re-establishment of medial cortical contact, use of a high-angle plate, and placement of the lag-screw in the center of the head as seen on the anteroposterior and lateral roentgenograms significantly increase the mean load to failure and decrease the rate of failure of fixation of unstable intertrochanteric fractures of the hip.
引用
收藏
页码:726 / 733
页数:8
相关论文
共 24 条
[1]  
CHANG WS, 1987, CLIN ORTHOP RELAT R, V225, P141
[2]  
Clark J M, 1987, J Arthroplasty, V2, P99, DOI 10.1016/S0883-5403(87)80016-5
[3]  
DANIS R, 1970, MANUAL INTERNAL FIXA, P10
[4]   UNSTABLE INTERTROCHANTERIC FRACTURES OF HIP [J].
DIMON, JH ;
HUGHSTON, JC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1967, A 49 (03) :440-&
[5]  
DOHERTY JH, 1979, CLIN ORTHOP RELAT R, V141, P184
[6]  
DOPPELT SH, 1980, ORTHOP CLIN N AM, V11, P507
[7]   TREATMENT OF TROCHANTERIC HIP FRACTURES USING A COMPRESSION SCREW [J].
ECKER, ML ;
JOYCE, JJ ;
KOHL, EJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1975, A 57 (01) :23-27
[8]   THE TREATMENT OF TROCHANTERIC FRACTURES OF THE FEMUR [J].
EVANS, EM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1949, 31 (02) :190-203
[9]  
HARRINGTON KD, 1973, SO MED J A, V55, P1367
[10]   TREATMENT OF INTERTROCHANTERIC FRACTURES OF THE FEMUR - A COMPARISON OF THE RICHARDS SCREW-PLATE WITH THE JEWETT NAIL-PLATE [J].
HEYSEMOORE, GH ;
MACEACHERN, AG ;
EVANS, DCJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1983, 65 (03) :262-267