DETERMINING RESIDUAL URINE VOLUMES USING A PORTABLE ULTRASONOGRAPHIC DEVICE

被引:26
作者
REVORD, JP
OPTIZ, JL
MURTAUGH, P
HARRISON, J
机构
[1] MAYO CLIN & MAYO FDN,DEPT PHYS MED & REHABIL,ROCHESTER,MN 55905
[2] MAYO CLIN & MAYO FDN,BIOSTAT SECT,ROCHESTER,MN 55905
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 1993年 / 74卷 / 05期
关键词
NEUROGENIC BLADDER; ULTRASONOGRAPHY;
D O I
10.1016/0003-9993(93)90104-I
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Residual urine volumes of 24 men with neurogenic bladder dysfunction were repetitively assessed 400 times with the BVI-2000 BladderScan portable ultrasonographic device prior to 100 episodes of intermittent catheterization. By comparing each examiner's first ultrasonographic measurement of urine volume with the catheterized urine volume, the mean error of the ultrasonographic measurements was -26 mL (-11%) and the mean absolute error was 44mL (22%). The ultrasonographic measurements detected the presence of residual urine volumes of ≥100mL with a sensitivity of 90% and a specificity of 81%. In the subset of catheterization episodes in which the catheterized urine volumes were ≤200mL, the mean error of the ultrasonographic measurements was -15mL (-9%), the mean absolute error was 37mL (28%), and the sensitivity and specificity were 77% and 81%, respectively. There was no clear advantage in using the average or maximum of two repeated ultrasonographic measurements over using each examiner's first ultrasonographic measurement alone. Increased examiner experience did not significantly decrease the erros encountered. © 1993.
引用
收藏
页码:457 / 462
页数:6
相关论文
共 27 条
  • [1] ALFTHAN O, 1969, Annales Chirurgiae et Gynaecologiae Fenniae, V58, P300
  • [2] ULTRASOUND ASSESSMENT OF RESIDUAL URINE - A QUANTITATIVE METHOD
    BEACOCK, CJM
    ROBERTS, EE
    REES, RWM
    BUCK, AC
    [J]. BRITISH JOURNAL OF UROLOGY, 1985, 57 (04): : 410 - 413
  • [3] CARDENAS DD, 1988, ARCH PHYS MED REHAB, V69, P514
  • [4] DADFAR H, 1972, INVEST UROL, V9, P363
  • [5] Doust B D, 1974, J Clin Ultrasound, V2, P127, DOI 10.1002/jcu.1870020209
  • [6] ACCURACY AND REPEATABILITY OF BLADDER VOLUME MEASUREMENT USING ULTRASONIC-IMAGING
    GRIFFITHS, CJ
    MURRAY, A
    RAMSDEN, PD
    [J]. JOURNAL OF UROLOGY, 1986, 136 (04) : 808 - 812
  • [7] AUSCULTATORY PERCUSSION OF THE URINARY-BLADDER
    GUARINO, JR
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1985, 145 (10) : 1823 - 1825
  • [8] THE ESTIMATION OF BLADDER VOLUME BY SONOCYSTOGRAPHY
    HAKENBERG, OW
    RYALL, RL
    LANGLOIS, SL
    MARSHALL, VR
    [J]. JOURNAL OF UROLOGY, 1983, 130 (02) : 249 - 251
  • [9] ULTRASOUND ASSESSMENT OF RESIDUAL URINE IN CHILDREN
    HARRISON, NW
    PARKS, C
    SHERWOOD, T
    [J]. BRITISH JOURNAL OF UROLOGY, 1975, 47 (07): : 805 - 814
  • [10] VERIFICATION OF THE ACCURACY AND RANGE OF TRANS-VAGINAL ULTRASOUND IN MEASURING BLADDER VOLUMES IN WOMEN
    HAYLEN, BT
    [J]. BRITISH JOURNAL OF UROLOGY, 1989, 64 (04): : 350 - 352