THE DETERMINATION OF PHANTOM AND COLLIMATOR SCATTER COMPONENTS OF THE OUTPUT OF MEGAVOLTAGE PHOTON BEAMS - MEASUREMENT OF THE COLLIMATOR SCATTER PART WITH A BEAM-COAXIAL NARROW CYLINDRICAL PHANTOM

被引:111
作者
VANGASTEREN, JJM
HEUKELOM, S
VANKLEFFENS, HJ
VANDERLAARSE, R
VENSELAAR, JLM
WESTERMANN, CF
机构
[1] CATHOLIC UNIV NIJMEGEN, DEPT RADIOTHERAPY, NIJMEGEN, NETHERLANDS
[2] NETHERLANDS CANC INST, ANTONI VAN LEEUWENHOEKHUIS, DEPT RADIOTHERAPY, 1066 CX AMSTERDAM, NETHERLANDS
[3] DR BERNARD VERBEETEN INST, TILBURG, NETHERLANDS
[4] WESTEINDE ZIEKENHUIS, DEPT CLIN PHYS, THE HAGUE, NETHERLANDS
[5] NUCLETRON RES, LEERSUM, NETHERLANDS
[6] DR DANIEL DEN HOED CANC CTR, ROTTERDAM, NETHERLANDS
关键词
PHANTOM SCATTER COMPONENT; COLLIMATOR SCATTER COMPONENT; MEGAVOLTAGE PHOTON BEAM;
D O I
10.1016/0167-8140(91)90124-Y
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The separation of the total scatter correction factor S(c),p in a collimator scatter component, S(c), and a phantom scatter component, S(p), has proven to be an useful concept in megavoltage photon beam dose calculations in situations which differ from the standard treatment geometry. A clinically applicable method to determine S(c) is described. Measurements are carried out with an ionization chamber, placed at a depth beyond the range of contaminant electrons, in a narrow cylindrical polystyrene phantom with a diameter of 4 cm of which the axis coincides with the beam axis. S(c),p is measured in a full-scatter phantom and S(p) can be derived from S(c),p and S(c). In order to obtain a reliable separation, i.e. excluding the influence of contaminant electrons, measurements of S(c),p have been carried out at depths of 5 cm for photon beams with a quality index (QI) up to and including 0.75 and a depth of 10 cm with QI larger than 0.75. These depths are in accordance with recommendations given in recent dosimetry protocols. The consistency of the method was checked by comparing calculated and measured values of S(c),p for a set of blocked fields for a range of photon beam energies from Co-60 up to 25 MV showing a maximum deviation of 2%. The method can easily be implemented in existing procedures for the calculation of the number of monitor units to deliver a specified dose to a target volume.