TEST-RESULTS FROM A POSITION-SENSITIVE NAI DETECTOR

被引:7
作者
BASSALLECK, B
HASINOFF, MD
SALOMON, M
机构
[1] Physics Dept., University of British Columbia, Vancouver
来源
NUCLEAR INSTRUMENTS & METHODS | 1979年 / 163卷 / 2-3期
关键词
D O I
10.1016/0029-554X(79)90123-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Results are reported for a position-sensitive NaI detector of 25.4 cm length and 5.1 cm thickness. Position resolutions as good as 1.0 cm fwhm for electrons and 1.4 cm fwhm for gamma rays have been obtained using left-right signals. Data for gamma rays and electrons of various energies are presented. © 1979.