CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .1. THE ATRIAL LEAD

被引:41
作者
GLIKSON, M
VONFELDT, LK
SUMAN, VJ
HAYES, DL
机构
[1] MAYO CLIN & MAYO FDN,DIV CARDIOVASC DIS & INTERNAL MED,200 1ST ST SW,ROCHESTER,MN 55905
[2] MAYO CLIN & MAYO FDN,BIOSTAT SECT,ROCHESTER,MN 55905
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 08期
关键词
PACING LEADS; ATRIAL SCREW-IN LEADS; LEAD PERFORMANCE; LEAD SURVIVAL;
D O I
10.1111/j.1540-8159.1994.tb02459.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Since 1989, 168 Telectronics model 330-801 active fixation, polyurethane insulated atrial leads (Accufix) have been implanted at the Mayo Clinic. There were four (2.4%) acute lead related complications, (i.e., perforation, microdislodgment, and pericarditis). Over a median follow-up time of 7.6 months (up to 2.7 years), there were 14 (8.3%) chronic complications, including 1 instance (0.6%) of definite lead failure. Most of these complications were early (within the first month) and transient. Four patients (2.4%) required reoperation for chronic complications. During follow-up, 23% of the examined patients had high pacing thresholds, most at about 3 months after implantation, necessitating high-output programming. The exact mechanism and natural history of this phenomenon should be further investigated.
引用
收藏
页码:1399 / 1404
页数:6
相关论文
共 13 条
  • [1] ANDERSON RL, 1991, PACE, V14, P738
  • [2] 3-YEAR CLINICAL-EXPERIENCE WITH A NEW ENDOCARDIAL SCREW-IN LEAD WITH INTRODUCTION PROTECTION FOR USE IN THE ATRIUM AND VENTRICLE
    BISPING, HJ
    KREUZER, J
    BIRKENHEIER, H
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1980, 3 (04): : 424 - 435
  • [3] Dodinot B., 1985, Pacemaker Leads. Proceedings of the International Symposium, P75
  • [4] FURMAN S, 1989, PRACTICE CARDIAC PAC, V2, P23
  • [5] A MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD
    HAYES, DL
    GRAHAM, KJ
    IRWIN, M
    VIDAILLET, H
    DISLER, G
    SWEESY, M
    OSBORN, MJ
    SUMAN, VJ
    NEUBAUER, SA
    SEEBANDT, M
    KALLINEN, L
    CROWSON, CS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (07): : 1033 - 1039
  • [6] ACTIVE FIXATION ATRIAL LEADS - RANDOMIZED COMPARISON OF 2 LEAD DESIGNS
    KAY, GN
    ANDERSON, K
    EPSTEIN, AE
    PLUMB, VJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (08): : 1355 - 1361
  • [7] KAYE G, 1991, EUR J CPE, V1, P154
  • [8] RELIABILITY OF ATRIAL SCREW-IN LEADS
    MARKEWITZ, A
    WENKE, K
    WEINHOLD, C
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11): : 1777 - 1783
  • [9] LONG-TERM FUNCTIONAL INTEGRITY OF ATRIAL LEADS
    PARSONNET, V
    HESSELSON, AB
    HARARI, DC
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (04): : 517 - 521
  • [10] PERRINS EJ, 1981, BRIT HEART J, V46, P196