PERFORMANCE OF A MATERIALS SCIENCE FACILITY AT THE PHOTON FACTORY - SOFT-X-RAY BEAMLINE BL 13C COUPLED WITH MULTIPURPOSE ANALYTICAL APPARATUS

被引:31
作者
SHIMADA, H [1 ]
MATSUBAYASHI, N [1 ]
IMAMURA, M [1 ]
SATO, T [1 ]
YOSHIMURA, Y [1 ]
HAYAKAWA, T [1 ]
TAKEHIRA, K [1 ]
TOYOSHIMA, A [1 ]
TANAKA, K [1 ]
NISHIJIMA, A [1 ]
机构
[1] NATL LAB HIGH ENERGY PHYS,PHOTON FACTORY,TSUKUBA,IBARAKI 305,JAPAN
关键词
D O I
10.1063/1.1145848
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The performance data of an undulator soft x-ray beamline based on the cylindrical element monochromator concept are described. Photon fluxes over 1012 photon/s 0.1% b.w. are achieved with the slit width of 100 μm at the harmonic peaks of a 13-period undulator. The maximum resolution obtained at the N K edge (E/ΔE) is ≈5000 with the slit width of 10 μm. The beamline is equipped with a multipurpose analytical instrument having a hemispherical electron analyzer and other detectors. The system is used for both x-ray photoelectron spectroscopy and x-ray absorption fine structure analyses for surface and materials science. © 1995 American Institute of Physics.
引用
收藏
页码:1780 / 1782
页数:3
相关论文
共 12 条
[1]   STUDY OF DIFFERENT SX-700 MONOCHROMATOR DESIGNS FOR THE UNDULATOR BEAMLINE (BL51) AT MAX-LAB [J].
AKSELA, S ;
KIVIMAKI, A ;
NYHOLM, R ;
SVENSSON, S .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :1252-1255
[2]   PERFORMANCE OF THE DRAGON SOFT-X-RAY BEAMLINE [J].
CHEN, CT ;
SETTE, F .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :1616-1621
[3]   CONCEPT AND DESIGN PROCEDURE FOR CYLINDRICAL ELEMENT MONOCHROMATORS FOR SYNCHROTRON RADIATION [J].
CHEN, CT .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 256 (03) :595-604
[4]   PHOTOELECTRIC QUANTUM EFFICIENCIES AND FILTER WINDOW ABSORPTION-COEFFICIENTS FROM 20-EV TO 10-KEV [J].
DAY, RH ;
LEE, P ;
SALOMAN, EB ;
NAGEL, DJ .
JOURNAL OF APPLIED PHYSICS, 1981, 52 (11) :6965-6973
[5]   STATUS AND PERSPECTIVES OF HIGH-RESOLUTION SPECTROSCOPY IN THE SOFT-X-RAY RANGE [J].
KAINDL, G ;
DOMKE, M ;
LAUBSCHAT, C ;
WESCHKE, E ;
XUE, C .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :1234-1240
[6]   INVESTIGATION OF ENERGY AND VIBRATIONAL STRUCTURE OF INNER SHELL (1S)-1(PI-2P)-PI-1 STATE OF NITROGEN MOLECULE BY ELECTRON-IMPACT WITH HIGH-RESOLUTION [J].
KING, GC ;
READ, FH ;
TRONC, M .
CHEMICAL PHYSICS LETTERS, 1977, 52 (01) :50-54
[7]   INSERTION DEVICES AT THE PHOTON FACTORY - PRESENT STATUS AND FUTURE-DEVELOPMENT [J].
KITAMURA, H ;
YAMAMOTO, S ;
KAMADA, S .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :1407-1413
[8]   COMBINED APPARATUS FOR X-RAY PHOTOELECTRON-SPECTROSCOPY AND EXTENDED X-RAY ABSORPTION FINE-STRUCTURE METHOD USING SYNCHROTRON RADIATION [J].
MATSUBAYASHI, N ;
KOJIMA, I ;
KURAHASHI, M ;
NISHIJIMA, A ;
ITOH, A ;
UTAKA, T .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :2533-2536
[9]   A SOFT-X-RAY BEAM LINE (BL13C) AT THE PHOTON FACTORY WITH A CEM USING UNDULATOR RADIATION [J].
MATSUBAYASHI, N ;
SHIMADA, H ;
TANAKA, K ;
SATO, T ;
YOSHIMURA, Y ;
NISHIJIMA, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :1363-1366
[10]   SOFT-X-RAY SPECTROSCOPY BEAM LINE ON THE NSLS X1 UNDULATOR - OPTICAL DESIGN AND 1ST PERFORMANCE TESTS [J].
RANDALL, KJ ;
FELDHAUS, J ;
ERLEBACH, W ;
BRADSHAW, AM ;
EBERHARDT, W ;
XU, Z ;
MA, Y ;
JOHNSON, PD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :1367-1370