INFRARED-TO-MILLIMETER, BROADBAND, SOLID STATE BOLOMETER DETECTORS

被引:5
作者
ALLEN, C
ARAMS, F
WANG, M
BRADLEY, CC
机构
[1] Airborne Instruments Laboratory, A Division of Cutler-Hammer, Melville, NY
[2] MIT National Magnet Laboratory, Cambridge, MA
来源
APPLIED OPTICS | 1969年 / 8卷 / 04期
关键词
D O I
10.1364/AO.8.000813
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Two versions of a thermal detector have been developed combining broadband response over the entire near infrared–to–millimeter region, room temperature operation, solid state construction, and amenability to absolute calibration. The sensitive detector version has a noise equivalent power (NEP) near 7 × 10−10 W/Hz1/2 with a time constant of 6 msec. The power measuring bolometer version has a sensitivity near 10−8 W for a 3-sec time constant. The basic construction uses a flake thermistor as a temperature sensor in combination with an optimized radiation absorber. © 1969 Optical Society of America.