EVALUATION AND TEST OF THE RESPONSE MATRIX OF A MULTISPHERE NEUTRON SPECTROMETER IN A WIDE ENERGY-RANGE .1. CALIBRATION

被引:17
作者
AROUA, A [1 ]
GRECESCU, M [1 ]
LERCH, P [1 ]
PRETRE, S [1 ]
VALLEY, JF [1 ]
VYLET, V [1 ]
机构
[1] SWISS NUCL SAFETY INSPECTORATE,CH-5232 VILLIGEN,SWITZERLAND
关键词
D O I
10.1016/0168-9002(92)90404-R
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A multisphere neutron spectrometer has been developed. It consists of a cylindrical He-3 thermal neutron counter and a set of 11 spherical moderators made of polyethylene with diameters ranging from 2 to 15 in. The spectrometer has been calibrated in two thermal and eight monoenergetic neutron beams (8.2 keV-14.8 MeV). The spectrometric system is described, and the results of the calibration are discussed. The thermal and high energy regions are well covered but there is a significant gap at intermediate energies. A comparison of the results with those of similar works is presented. The differences obtained underline the necessity for a proper specific calibration for each multisphere spectrometric system.
引用
收藏
页码:298 / 304
页数:7
相关论文
共 26 条
[1]  
ALBERTS WG, 1986, EUR8486 COMM EUR COM, P176
[2]  
ALEVRA AV, 1988, RADIAT PROT DOSIM, V23, P293
[3]  
ALEVRA AV, 1986, PTBND26 REP
[4]  
AROUA A, 1990, EVALUATION MESURES S
[5]   APPLICATIONS OF BONNER SPHERE DETECTORS IN NEUTRON FIELD DOSIMETRY [J].
AWSCHALOM, M ;
SANNA, RS .
RADIATION PROTECTION DOSIMETRY, 1985, 10 (1-4) :89-101
[6]   A NEW TYPE OF NEUTRON SPECTROMETER [J].
BRAMBLETT, RL ;
EWING, RI ;
BONNER, TW .
NUCLEAR INSTRUMENTS & METHODS, 1960, 9 (01) :1-12
[7]  
CAIZERGUES R, 1972, CEAR4400
[8]  
CROSS WG, 1987, DOSIMETRY IONIZING R, V2, pCH2
[9]  
DOLIAS M, 1972, EUR4791F COMM EUR CO
[10]  
HANKINS DE, 1968, HEALTH PHYS, V14, P518