ESOPHAGEAL ELECTRODES ALLOW PRECISE ASSESSMENT OF CARDIAC-OUTPUT BY BIOIMPEDANCE

被引:16
作者
BALESTRA, B
MALACRIDA, R
LEONARDI, L
SUTER, P
MARONE, C
机构
[1] OSPED SAN GIOVANNI BELLINZONA, CTR CURE INTENSE, CH-6500 BELLINZONA, SWITZERLAND
[2] UNIV GENEVA, DEPT ANAESTHESIOL, CH-1211 GENEVA 4, SWITZERLAND
关键词
ELECTRODES; CARDIAC OUTPUT; THERMODILUTION; EXERCISE; HEMODYNAMICS; CARDIOGRAPHY; IMPEDANCE; INTUBATION; ESOPHAGEAL; CRITICAL CARE; INTENSIVE CARE UNIT; CATHETERIZATION; PULMONARY ARTERY;
D O I
10.1097/00003246-199201000-00018
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To analyze the impact of the position of the thoracic external electrodes on the values of cardiac output measured by electrical bioimpedance and to compare the results obtained by bioimpedance with those values determined by thermodilution in critically ill patients. Design: Open, prospective, comparative trial. Setting: ICU of a teaching hospital. Patients: Twenty healthy volunteers and ten critically ill patients. Interventions: Measurements of cardiac output by bioimpedance at rest and after physical activity in normal volunteers and after changing the neck or xiphoid electrodes. Comparisons of cardiac output obtained by thermodilution and bioimpedance with internal and external electrodes in patients. Measurements and Main Results: Mean +/- SD values are presented. Cardiac output values at rest and after exercise were 6.7 +/- 1.3 and 10.8 +/- 2.6 L/min at rest and after exercise, respectively (p < .001). Displacement of the xiphoid electrodes 3 cm in the caudal direction was accompanied by a decrease of the mean cardiac output from 7.1 +/- 1.2 to 5.8 +/- 1.3 L/min (p < .001) and displacement 3 and 6 cm cranially was accompanied by increases in cardiac output from 7.1 +/- 1.2 to 8.1 +/- 1.4 L/min (p < .001) and 8.6 +/- 1.5 L/min (p < .001), respectively. In the ten patients, cardiac output measurements were virtually identical when results obtained by thermodilution (6.7 +/- 3.1 L/min) were compared with those results obtained by bioimpedance using internal esophageal (6.6 +/- 3.1 L/min), but not external (4.7 +/- 1.6 L/min) electrodes. Conclusions: a) The values of cardiac output derived from measurements obtained by bioimpedance using internal electrodes were comparable with those values derived from thermodilution.b) Values of cardiac output from bioimpedance studies with external electrodes were dependent on the position of the xiphoid electrodes.
引用
收藏
页码:62 / 67
页数:6
相关论文
共 42 条
[1]   COMPARISON OF MEASUREMENTS OF CARDIAC-OUTPUT BY BIOIMPEDANCE AND THERMODILUTION IN SEVERELY ILL SURGICAL PATIENTS [J].
APPEL, PL ;
KRAM, HB ;
MACKABEE, J ;
FLEMING, AW ;
SHOEMAKER, WC .
CRITICAL CARE MEDICINE, 1986, 14 (11) :933-935
[2]  
BACHE RJ, 1969, AM J MED SCI, V258, P100
[3]  
BAKER LE, 1971, CARDIOVASC RES CENT, V94, P135
[4]   CONTINUOUS NONINVASIVE REAL-TIME MONITORING OF STROKE VOLUME AND CARDIAC-OUTPUT BY THORACIC ELECTRICAL BIOIMPEDANCE [J].
BERNSTEIN, DP .
CRITICAL CARE MEDICINE, 1986, 14 (10) :898-901
[5]   A NEW STROKE VOLUME EQUATION FOR THORACIC ELECTRICAL BIOIMPEDANCE - THEORY AND RATIONALE [J].
BERNSTEIN, DP .
CRITICAL CARE MEDICINE, 1986, 14 (10) :904-909
[6]   IMPEDANCE CARDIOGRAPHY, A METHOD TO EVALUATE QUANTITATIVELY CARDIAC-OUTPUT - COMPARISON WITH FICK PRINCIPLE [J].
BETZ, R ;
BASTANIER, CK ;
MOCELLIN, R .
BASIC RESEARCH IN CARDIOLOGY, 1977, 72 (01) :46-56
[7]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[8]  
BOLDT J, 1988, ANAESTHESIST, V37, P218
[9]  
DELFINO M, 1984, Giornale Italiano di Cardiologia, V14, P101
[10]   MEASUREMENT OF CARDIAC-OUTPUT BY ELECTRICAL IMPEDANCE AT REST AND DURING EXERCISE [J].
DENNISTON, JC ;
MAHER, JT ;
REEVES, JT ;
CRUZ, JC ;
CYMERMAN, A ;
GROVER, RF .
JOURNAL OF APPLIED PHYSIOLOGY, 1976, 40 (01) :91-95